文献速递神经介入CTDSA

高分辨率C臂CT联合条状金属伪影去除技术在颅内动脉瘤支架辅助栓塞中的应用

Application of High-Resolution C-Arm CT Combined with Streak Metal Artifact Removal Technology for the Stent-Assisted Embolization of Intracranial Aneurysms.
PubMed 文献 · AJNR Am J Neuroradiol · 2019-09-26

2017年10月至2018年7月郑州大学第一附属医院对107例颅内动脉瘤患者(共118个动脉瘤)行支架辅助栓塞,术中及术后行常规与高分辨率C臂CT扫描并采用条状金属伪影去除技术重建,结果显示高分辨率C臂CT联合条状金属伪影去除重建的图像质量显著更高(P < .05),可显示支架展开程度与栓塞密度。

要点

  • 高分辨率C臂CT联合条状金属伪影去除重建的图像质量显著高于常规重建(P < .05)。
  • 条状金属伪影去除重建联合工作角度2D血管造影显示6枚支架展开不全和15例患者动脉瘤栓塞不全,随后均得到解决。
  • 术后随访93例患者6-13个月,发现3例动脉瘤复发。

核心结论

  • 高分辨率C臂CT联合条状金属伪影去除技术有效减少了动脉瘤栓塞过程中支架和弹簧圈的金属伪影。
  • 该方法可帮助医师判断支架展开程度和弹簧圈栓塞密度,从而可能减少并发症和动脉瘤复发。

研究设计与样本

  • 研究类型为单中心临床研究,2017年10月至2018年7月郑州大学第一附属医院治疗107例颅内动脉瘤患者(共118个动脉瘤),均行支架辅助栓塞。
  • 术中和术后对支架区域行常规C臂CT和高分辨率C臂CT扫描,分别采用和不采用条状金属伪影去除技术重建3D图像,并比较图像质量。
  • 重建图像用于判断支架展开程度和栓塞密度;随访评估包括临床和血管造影结果及并发症。

临床应用与阅片要点

  • 高分辨率C臂CT联合条状金属伪影去除重建图像质量显著更高(P < .05),可用于判断支架展开程度和弹簧圈栓塞密度。
  • 条状金属伪影去除重建联合工作角度2D血管造影可发现支架展开不全(6枚支架)和动脉瘤栓塞不全(15例患者),并随后得到解决。
  • 术后1例出血,随访93例患者6-13个月发现3例动脉瘤复发。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: Metal artifacts from coils and stents limit the level of detail in C-arm CT images of stent attachment and coiling attenuation in the aneurysm neck. We evaluated the utility of high-resolution C-arm CT combined with streak metal artifact removal technology for stent-assisted embolization of intracranial aneurysms.

MATERIALS AND METHODS: From October 2017 to July 2018, the First Affiliated Hospital of Zhengzhou University treated 107 patients with intracranial aneurysms (118 aneurysms in total) with stent-assisted embolization. Conventional C-arm CT and high-resolution C-arm CT scanning of the stented area were performed during and after treatment. 3D images were reconstructed with and without streak metal artifact removal techniques. Subsequently, the image quality was compared. The reconstructed images indicated the stent deployment degree and packing density. Follow-up assessments included clinical and angiographic outcomes and complications.

RESULTS: In total, 118 aneurysms were successfully embolized using 118 stents. Image quality was significantly higher (P < .05) with high-resolution C-arm CT combined with streak metal artifact removal reconstruction. Streak metal artifact removal reconstruction and 2D angiography at working angles showed incomplete deployment of 6 stents and incomplete aneurysm embolization of 15 patients, which were subsequently resolved. One case of hemorrhage was noted postoperatively. Follow-up of 93 patients at 6-13 months indicated 3 cases of aneurysm recurrence.

CONCLUSIONS: High-resolution C-arm CT combined with the streak metal artifact removal technique effectively reduced metal artifacts from stents and coils during aneurysm embolization. This method can help physicians determine the extent of stent deployment and the packing density of coils and thus potentially reduce complications and aneurysm recurrence.

来源:AJNR Am J Neuroradiol,查看原文。版权归原作者所有。

引用

  1. Li TF, Ma J, Han XW, et al. Application of High-Resolution C-Arm CT Combined with Streak Metal Artifact Removal Technology for the Stent-Assisted Embolization of Intracranial Aneurysms. AJNR Am J Neuroradiol. 2019. PMID: 31558505. doi:10.3174/ajnr.A6190

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。