脑静脉血栓形成:增强血液信号抑制的MR黑血血栓成像

Cerebral Venous Thrombosis: MR Black-Blood Thrombus Imaging with Enhanced Blood Signal Suppression.
PubMed 文献 · AJNR Am J Neuroradiol · 2019-09-26

本研究纳入26名参与者(13名志愿者和13名患者),比较开启与关闭DANTE准备的黑血血栓成像技术。结果显示DANTE准备可显著降低残余血信号及其与正常管腔的对比噪声比,并提高血栓与残余血的对比噪声比、诊断特异度及与传统影像方法的一致性。作者结论认为DANTE准备的黑血血栓成像技术是诊断脑静脉窦血栓形成的可靠MR成像技术。

要点

  • DANTE准备的黑血血栓成像可显著抑制残余血信号,减少血流伪影对血栓判读的干扰。
  • 该技术提高血栓与残余血的对比噪声比,并提升诊断特异度及与传统影像方法的一致性。
  • 作者认为DANTE准备的黑血血栓成像是诊断脑静脉窦血栓形成的可靠MR技术。

核心结论

  • DANTE准备的黑血血栓成像是诊断脑静脉窦血栓形成的可靠MR成像技术。

研究设计与样本

  • 研究纳入26名参与者(13名志愿者和13名患者),均接受传统影像方法检查后随机进行2次黑血血栓成像扫描,分别开启和关闭预先优化的DANTE准备。

关键影像表现/征象

  • DANTE准备的黑血血栓成像上残余血信号被显著抑制,残余血与正常管腔的对比噪声比显著降低。
  • DANTE准备的黑血血栓成像提供更高的血栓与残余血对比噪声比。
  • DANTE准备的黑血血栓成像测得的血栓体积显著更低。

诊断标准与关键数值

  • 志愿者组DANTE准备开启 vs 关闭的残余血SNR为11.3 ± 2.9 vs 54.0 ± 23.4,P < .001,DANTE准备开启时降低。
  • 志愿者组DANTE准备开启 vs 关闭的残余血与正常管腔CNR为7.5 ± 3.4 vs 49.2 ± 23.3,P < .001,DANTE准备开启时降低。
  • 患者组DANTE准备开启 vs 关闭的残余血SNR为16.4 ± 8.0 vs 75.0 ± 35.1,P = .002,DANTE准备开启时降低。
  • 患者组DANTE准备开启 vs 关闭的残余血与正常管腔CNR为12.4 ± 7.8 vs 68.8 ± 35.4,P = .002,DANTE准备开启时降低。
  • 以传统影像技术联合作为参考,DANTE准备的黑血血栓成像诊断特异度及与传统影像方法的一致性均 substantially improved。

临床应用与阅片要点

  • 作者认为DANTE准备的黑血血栓成像是诊断脑静脉窦血栓形成的可靠MR成像技术。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: The residual blood flow artifact is a critical confounder for MR black-blood thrombus imaging of cerebral venous sinus thrombosis. This study aimed to conduct a validation of a new MR black-blood thrombus imaging technique with enhanced blood signal suppression.

MATERIALS AND METHODS: Twenty-six participants (13 volunteers and 13 patients) underwent conventional imaging methods followed by 2 randomized black-blood thrombus imaging scans, with a preoptimized delay alternating with nutation for tailored excitation (DANTE) preparation switched on and off, respectively. The signal intensity of residual blood, thrombus, brain parenchyma, normal lumen, and noise on black-blood thrombus images were measured. The thrombus volume, SNR of residual blood, and contrast-to-noise ratio for residual blood versus normal lumen, thrombus versus residual blood, and brain parenchyma versus normal lumen were compared between the 2 black-blood thrombus imaging techniques. Segmental diagnosis of venous sinus thrombosis was evaluated for each black-blood thrombus imaging technique using a combination of conventional imaging techniques as a reference.

RESULTS: In the volunteer group, the SNR of residual blood (11.3 ± 2.9 versus 54.0 ± 23.4, P < .001) and residual blood-to-normal lumen contrast-to-noise ratio (7.5 ± 3.4 versus 49.2 ± 23.3, P < .001) were significantly reduced using the DANTE preparation. In the patient group, the SNR of residual blood (16.4 ± 8.0 versus 75.0 ± 35.1, P = .002) and residual blood-to-normal lumen contrast-to-noise ratio (12.4 ± 7.8 versus 68.8 ± 35.4, P = .002) were also significantly lower on DANTE-prepared black-blood thrombus imaging. The new black-blood thrombus imaging technique provided higher thrombus-to-residual blood contrast-to-noise ratio, significantly lower thrombus volume, and substantially improved diagnostic specificity and agreement with conventional imaging methods.

CONCLUSIONS: DANTE-prepared black-blood thrombus imaging is a reliable MR imaging technique for diagnosing cerebral venous sinus thrombosis.

来源:AJNR Am J Neuroradiol,查看原文。版权归原作者所有。

引用

  1. Wang G, Yang X, Duan J, et al. Cerebral Venous Thrombosis: MR Black-Blood Thrombus Imaging with Enhanced Blood Signal Suppression. AJNR Am J Neuroradiol. 2019. PMID: 31558501. doi:10.3174/ajnr.A6212

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。