剪切波弹性成像与儿童肝病患者肝纤维化评分相关

Shear wave elastography correlates with liver fibrosis scores in pediatric patients with liver disease.
PubMed 文献 · Pediatr Radiol · 2019-08-15

纳入0-18岁已知或疑似肝病且30天内行肝活检的患儿70例,与对照组79例比较,采用GE LOGIQ E9系统行二维剪切波弹性成像(2-D SWE)。2-D SWE与METAVIR及Ishak纤维化评分中度相关(rs=0.43),识别F1期纤维化的AUC为0.89(95%CI 0.83-0.95),敏感度94.6%、特异度78.6%,理想截断值为1.29 m/s。

要点

  • 2-D SWE与METAVIR及Ishak纤维化评分均呈中度相关(rs=0.43)。
  • 识别F1期纤维化(对照/F0合并)的AUC为0.89,敏感度94.6%、特异度78.6%。
  • 识别纤维化的理想截断值为1.29 m/s,2-D SWE更擅长识别早期纤维化。

核心结论

  • 肝脏2-D SWE测量值与肝纤维化组织学评分相关,不受组织病理评分系统影响,但2-D SWE更擅长识别早期纤维化患者,而非区分各纤维化级别。使用GE LOGIQ超声系统的2-D SWE有助于识别有肝纤维化风险的儿科患者。

研究设计与样本

  • 研究纳入0-18岁已知或疑似肝病且在30天内行肝活检的受试者70例,并与对照组79例比较;使用GE LOGIQ E9系统行二维剪切波弹性成像(2-D SWE);肝活检标本按METAVIR和Ishak评分系统评分,采用Spearman's Rho相关、ROC分析、Kruskal-Wallis检验和Mann-Whitney U检验。

诊断标准与关键数值

  • METAVIR F0和Ishak 0分者的2-D SWE中位数(1.35 m/s;1.36 m/s)显著低于更晚期肝病者(F1-F3:1.49-1.62 m/s;Ishak 1-4:1.45-1.63 m/s),P<0.05。
  • F1期与对照/F0合并比较的AUC为0.89(95%CI 0.83-0.95;P<0.001),敏感度94.6%,特异度78.6%;Ishak 1分结果相似。
  • 识别纤维化的理想截断值为1.29 m/s。
  • 纤维化评分与2-D SWE中位数测量值中度相关(rs=0.43)。

临床应用与阅片要点

  • 作者结论认为,使用GE LOGIQ超声系统的2-D SWE有助于识别有肝纤维化风险的儿科患者;2-D SWE更擅长识别早期纤维化患者,而非区分各纤维化级别。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: While liver biopsy remains the gold standard, given the procedure risks and sampling errors, there is a need for reliable noninvasive biomarkers of hepatic fibrosis.

OBJECTIVE: Determine the accuracy of two-dimensional shear wave elastography (2-D SWE) in predicting the histological severity of liver fibrosis in pediatric patients with known or suspected liver disease.

MATERIALS AND METHODS: Subjects 0-18 years old with known or suspected liver disease and liver biopsy within 30 days (n=70) were included. Comparisons by 2-D SWE were made to a control group (n=79). Two-dimensional SWE was performed using the GE LOGIQ E9 system. Liver biopsy specimens were scored according to METAVIR and Ishak scoring systems using Spearman's Rho correlation. Receiver operator characteristic (ROC) analysis, Kruskal-Wallis and Mann-Whitney U tests were conducted.

RESULTS: Control group median 2-D SWE measurements were lower than in subjects with any degree of liver fibrosis (P<0.001). Those with METAVIR F0 and Ishak 0 scores had significantly lower median 2-D SWE measurements (1.35 m/s; 1.36 m/s) than those with more advanced liver disease (F1-F3: 1.49-1.62 m/s; 1-4: 1.45-1.63 m/s) (P<0.05 for all), whereas the 2-D SWE in the higher scores were similar. Results did not differ between METAVIR and Ishak scores for any degree of fibrosis. Fibrosis scores moderately correlated with median 2-D SWE measurements (rs=0.43). The area under the curve for F1 compared to combined control/F0 was 0.89 (95% confidence interval [CI] 0.83-0.95; P<0.001) with sensitivity of 94.6% and specificity of 78.6%. Results for Ishak score 1 were similar. The ideal cutoff value for identifying fibrosis was determined to be 1.29 m/s.

CONCLUSION: The liver 2-D SWE measurements correlated with the histological liver fibrosis scores, regardless of the histopathological scoring system, although 2-D SWE was better at identifying patients with early fibrosis, not at distinguishing among the individual fibrosis levels. Two-dimensional SWE using the GE LOGIQ US system is useful for identifying pediatric patients at risk for liver fibrosis.

来源:Pediatr Radiol,查看原文。版权归原作者所有。

引用

  1. Farmakis SG, Buchanan PM, Guzman MA, et al. Shear wave elastography correlates with liver fibrosis scores in pediatric patients with liver disease. Pediatr Radiol. 2019. PMID: 31418057. doi:10.1007/s00247-019-04493-3

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。