基于结构与流体分析的现场CT衍生血流储备分数测量的操作者间与操作者内变异性:综合分析
本回顾性分析纳入22例同时接受320排冠状动脉CT血管造影和导管衍生血流储备分数(FFR)检查的症状性患者,13名不同经验水平的操作者经两种培训方式后测量24支血管的CT FFR。操作者间总体ICC为0.71,面对面培训组操作者间差异小于自学组(0.023±0.018 vs 0.031±0.024;P=0.024),CT FFR与导管FFR相关系数r=0.61。
要点
- 不同经验操作者现场测量CT FFR的总体重复性良好,操作者间ICC为0.71。
- 面对面培训组操作者间差异小于自学组(0.023±0.018 vs 0.031±0.024;P=0.024)。
- CT FFR与导管FFR相关系数r=0.61,平均绝对差0.096±0.089。
核心结论
- 基于结构与流体分析现场计算的CT FFR在不同经验操作者间观察到良好重复性,面对面培训可能减少变异性。
研究设计与样本
- 回顾性分析前瞻性入组队列,纳入22例症状性患者,均于90天内接受320排冠状动脉CT血管造影和导管衍生FFR检查,13名不同经验操作者分为两组培训后测量22例患者24支血管的CT FFR。
诊断标准与关键数值
- 操作者间总体ICC为0.71(95%CI:0.58,0.83),平均绝对差为0.027±0.022;第2组操作者间差异大于第1组(0.031±0.024 vs 0.023±0.018;P=0.024);重复计算者首次与第二次CT FFR均值无显著差异(ICC=0.66;95%CI:0.46,0.87),医学专家操作者内变异性最低(0.053±0.060);CT FFR与导管FFR总体相关系数r=0.61,平均绝对差0.096±0.089。
临床应用与阅片要点
- 作者认为面对面培训课程可能减少变异性。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To measure the inter- and intraobserver variability among operators of varying expertise in conducting CT-derived fractional flow reserve (CT FFR) measurements on-site by using structural and fluid analysis and to evaluate differences in reproducibility between two different training methods for end users.
MATERIALS AND METHODS: This retrospective analysis of the prospectively enrolled cohort included 22 symptomatic patients who underwent both 320-detector row coronary CT angiography and catheter-derived fractional flow reserve (FFR) within 90 days. Thirteen operators of varying expertise were assigned to one of two training arms: arm 1, on-site training by a specialist in CT FFR technology; arm 2, self-training through use of written materials. After the training, all 13 operators reviewed the CT data and measured CT FFR in 24 vessels in 22 patients. Inter- and intraoperator variability and agreements between CT FFR and catheter-derived FFR measurements were evaluated.
RESULTS: The overall intraclass correlation coefficient (ICC) among operators was 0.71 (95% confidence interval: 0.58, 0.83) with a mean absolute difference (± standard deviation) of 0.027 ± 0.022. The operators in arm 2 showed greater interoperator differences than those in arm 1 (0.031 ± 0.024 vs 0.023 ± 0.018; P = .024). Among operators who recalculated CT FFR, the mean CT FFR value did not significantly differ between the first and second calculations (ICC, 0.66; 95% confidence interval: 0.46, 0.87), with the medical specialists producing the lowest intraoperator variability (0.053 ± 0.060). The overall correlation coefficient between CT FFR and catheter FFR was r = 0.61, with a mean absolute difference of 0.096 ± 0.089.
CONCLUSION: Good reproducibility of CT FFR values calculated on-site on the basis of structural and fluid analysis was observed among operators of varying expertise. Face-to-face training sessions may cause less variability.© RSNA, 2019Supplemental material is available for this article.
引用
- Kumamaru KK, Angel E, Sommer KN, et al. Inter- and Intraoperator Variability in Measurement of On-Site CT-derived Fractional Flow Reserve Based on Structural and Fluid Analysis: A Comprehensive Analysis. Radiol Cardiothorac Imaging. 2019. PMID: 33778507. doi:10.1148/ryct.2019180012
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。