利用心外膜微型换能器估测区域心肌收缩
本文介绍一种将两个直径3 mm、10 MHz超声换能器缝在心外膜表面、以心外膜为参考补偿心脏运动、在猪模型上以2500 s-1脉冲重复频率记录射频信号来估测径向组织速度、应变率与应变并追踪心肌分层运动的超声系统。
要点
- 心外膜微型换能器以心外膜为参考可补偿心脏整体运动,短距离支持高频与高脉冲重复频率。
- 系统在猪模型上可估测心肌速度与应变模式,并追踪心肌分层运动。
- 心肌按四层划分,心内膜边界采用基于模糊逻辑并滤波去噪、剔除离群点的算法检测。
核心结论
- 猪模型结果证明该系统能够估测心肌速度与应变模式并追踪心肌分层运动,从而获得区域心肌功能的详细信息。
研究设计与样本
- 本文为技术性论文,详细描述系统设计、硬件电子学与算法,并给出处理后的速度模式与心肌应变曲线示例。
- 在猪模型上测试:两个直径3 mm、10 MHz超声换能器缝于心外膜表面。
- 系统以脉冲回波模式工作,使用为本应用开发的电子学,射频线以2500 s-1的脉冲重复频率记录。
- 以心外膜表面作为速度与应变估测的参考,该过程补偿了心脏的运动。
- 心内膜边界采用基于模糊逻辑并滤波去噪、剔除离群点的算法检测,心肌被分为四层。
关键影像表现/征象
- 由记录的射频信号计算心肌内径向组织速度随深度的变化,并据此估测径向应变率与应变、追踪心肌分层运动。
- 文中给出处理后的速度模式与心肌应变曲线示例。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
This paper describes an ultrasound system to monitor cardiac motion using miniature transducers attached directly to the epicardial surface. Our aim was to develop both a research tool for detailed studies of cardiac mechanics and a continuous, real time system for peri-operative evaluation of heart function. The system was tested on a porcine model. Two 3 mm diameter, 10 MHz ultrasound transducers were sutured to the epicardial surface. As the epicardial surface was the reference for the velocity and strain estimations, this procedure compensated for the motion of the heart. The short distance allowed for the use of high frequencies and pulse repetition rates. The system was driven in pulse-echo mode, using electronics developed for the application, and radio frequency (RF) lines were recorded at a pulse repetition rate of 2500 s-1. The endocardial border was detected using an algorithm based on fuzzy logic with filtration to reduce noise and remove outliers, and the myocardium was divided into four layers. Inside the myocardium, radial tissue velocity as a function of depth was calculated from the recorded RF signals, and the velocity estimates were used to estimate radial strain rate and strain and to track the motion of the myocardial layers. The scope of this paper is technical, giving a detailed description of system design, hardware electronics and algorithms, with examples of processed velocity patterns and myocardial strain curves. The results from this study on a porcine model demonstrate the system's ability to estimate myocardial velocity and strain patterns and to track the motion of the myocardial layers, thereby obtaining detailed information of the regional function of the myocardium.
引用
- Nguyen TT, Espinoza AW, Hyler S, et al. Estimating Regional Myocardial Contraction Using Miniature Transducers on the Epicardium. Ultrasound Med Biol. 2019. PMID: 31447239. doi:10.1016/j.ultrasmedbio.2019.07.416
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。