计算机断层扫描在经导管三尖瓣开发中的应用
本研究分析195例心脏舒张末期CT,比较严重三尖瓣反流患者与无三尖瓣反流患者的解剖差异,评估经导管三尖瓣置换的挑战及静脉入路决定因素。结果显示严重三尖瓣反流患者三尖瓣环更大、更圆,经颈静脉入路较经股静脉更直且个体差异更小。
要点
- 严重三尖瓣反流患者三尖瓣环更大、更圆,瓣环面积及隔侧-外侧、前后径均大于无三尖瓣反流者。
- 三尖瓣环至下腔静脉角度较至上腔静脉更锐利且个体差异大(均值101±18°,范围44°至164°)。
- 三尖瓣环至冠状静脉窦、右室流出道及调节束距离分别为11.4±3.8 mm、17.2±3.4 mm、31.0±6.7 mm,组间无差异。
核心结论
- 经颈静脉入路行经导管三尖瓣置换较经股静脉入路更直且重复性更好;假体开发可能受冠状静脉窦靠近、调节束存在及严重三尖瓣反流患者三尖瓣环较大等因素挑战。
研究设计与样本
- 研究分析195例心脏舒张末期CT,患者分别因严重三尖瓣反流(n=38)、经导管主动脉瓣置换(n=89)或左心耳封堵(n=68)术前接受检查;将有严重三尖瓣反流患者(n=68,其中原发性19例、继发性49例)与无三尖瓣反流患者(n=127)进行比较。
关键影像表现/征象
- 严重三尖瓣反流患者三尖瓣环更圆,偏心指数为0.88±0.08,P<0.001。
- 严重三尖瓣反流患者三尖瓣环面积更大,P<0.0001。
- 严重三尖瓣反流患者三尖瓣环隔侧-外侧径为45.3±8.0 mm,P<0.0001,前后径为44.4±7.4 mm,P<0.0001,均大于无三尖瓣反流患者。
- 三尖瓣环至下腔静脉角度均值为101±18°,范围44°至164°,较至上腔静脉角度(均值143±9°)更锐利且范围更宽。
- 三尖瓣环至冠状静脉窦、右室流出道及调节束距离分别为11.4±3.8 mm、17.2±3.4 mm、31.0±6.7 mm,组间无差异。
诊断标准与关键数值
- 严重三尖瓣反流患者三尖瓣环偏心指数为0.88±0.08(P<0.001),三尖瓣环隔侧-外侧径45.3±8.0 mm(P<0.0001),前后径44.4±7.4 mm(P<0.0001)。
- 三尖瓣环至下腔静脉角度均值101±18°(范围44°至164°),至冠状静脉窦距离11.4±3.8 mm,至右室流出道距离17.2±3.4 mm,至调节束距离31.0±6.7 mm。
临床应用与阅片要点
- 经颈静脉入路可能为到达三尖瓣环的首选入路,较经股静脉入路更直且个体间变异更小。
- 假体开发需考虑冠状静脉窦开口靠近三尖瓣环,可能需要短心房长度的假体。
- 严重三尖瓣反流患者三尖瓣环较大,可能对假体尺寸选择构成挑战。
- 调节束的存在可能对假体植入构成挑战。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Patients with severe symptomatic tricuspid regurgitation (TR) are often deemed ineligible for surgical valve repair due to comorbidities. In this setting, transcatheter tricuspid valve replacement (TTVR) is undergoing development, but delivery technique and prosthesis design have yet to be optimized. We sought to assess the challenges of TTVR and the determinants of venous route using computed tomography (CT) analysis.
METHODS AND RESULTS: A total of 195 end-diastolic cardiac CT performed prior to surgical correction of a severe TR (n = 38), transcatheter aortic valve replacement (n = 89), or left atrial appendage closure (n = 68) were analyzed. Patients with TR (n = 68; 19 primary and 49 secondary) were compared with patients without (n = 127). Continuous variables with normal and non-normal distributions were compared using Student t test or Mann-Whitney test respectively. The angle from the tricuspid annulus (TA) to the inferior vena cava was tighter (mean = 101 ± 18°) with a broader range of value (44° to 164°) than to the superior vena cava (mean = 143 ± 9°). Patients with TR had rounder TA (eccentricity index of 0.88 ± 0.08, p < 0.001), with a larger area (p < 0.0001), and septolateral (45.3 ± 8.0 mm, p < 0.0001) and anteroposterior (44.4 ± 7.4 mm, p < 0.0001) diameters than patients without. The distances from the TA to the coronary sinus, the right ventricular outflow tract, and the moderator band were respectively 11.4 ± 3.8 mm, 17.2 ± 3.4 mm, and 31.0 ± 6.7 mm, without differences between groups.
CONCLUSION: The transjugular access for TTVR is straighter and more reproducible than the transfemoral access. Prosthesis development may be challenged by the close position of the coronary sinus, the presence of a moderator band, and the large TA size of patients with severe TR.
KEY POINTS: • The tricuspid annulus is larger in patients with severe tricuspid regurgitation, confirming existing data. • The coronary sinus ostium is close to the tricuspid annulus, requiring a prosthesis with a short atrial length. • The transjugular venous route may be the preferred access to the tricuspid annulus, straighter with less inter-individual variations than the transfemoral route.
引用
- Chiaroni PM, Ternacle J, Deux JF, et al. Computed tomography for transcatheter tricuspid valve development. Eur Radiol. 2019. PMID: 31451974. doi:10.1007/s00330-019-06410-6
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。