胸部快速kV切换与双源双能CT技术图像质量和辐射剂量的比较

Comparison of image quality and radiation doses between rapid kV-switching and dual-source DECT techniques in the chest.
PubMed 文献 · Eur J Radiol · 2019-08-13

回顾性纳入97例患者,均行快速kV切换单源双能CT(SS-DECT)和双源双能CT(DS-DECT)胸部增强扫描,比较图像质量与辐射剂量。两种技术患者体重和体型特异性剂量估计(SSDE)无显著差异,虚拟单能图像上病灶和伪影评估无差异;但SS-DECT在63-71/97例患者材料分解碘图(MDI)上出现明显肺灌注异质性伪影,而DS-DECT无此表现(p < 0.001)。

要点

  • SS-DECT的MDI图像上63-71/97例出现肺灌注异质性伪影,可混淆真实肺灌注评估。
  • 两种技术虚拟单能图像上病灶和伪影评估无显著差异,SSDE无显著差异。
  • 阅读SS-DECT的MDI图像时需警惕肺灌注低灌注区可能为伪影。

核心结论

  • 尽管患者和辐射剂量相同,SS-DECT与DS-DECT生成的材料分解碘图存在显著差异;SS-DECT的MDI图像上肺灌注异质性导致低灌注伪影区,可混淆真实肺灌注的判读。

研究设计与样本

  • 回顾性研究,经机构审查委员会批准,纳入97例患者(男54例,女43例;年龄63 ± 14岁),均按科室标准方案行胸部增强双能CT,分别使用快速kV切换单源(SS-DECT)和双源(DS-DECT)技术;两种扫描仪重建的虚拟单能图像和材料分解碘(MDI)图像层厚和层间隔一致;两名胸部放射科医师独立评估所有双能CT的病变、图像质量、灌注缺损(MDI)和伪影;记录辐射剂量描述符体型特异性剂量估计(SSDE);数据分析采用Wilcoxon符号秩检验和Cohen's Kappa检验。

关键影像表现/征象

  • SS-DECT在63-71/97例患者的MDI图像上出现明显的肺灌注异质性伪影,而DS-DECT无此表现(p < 0.001)。
  • 两种技术虚拟单能图像上病变和伪影评估无差异(p > 0.05)。
  • 两种技术患者体重和SSDE无显著差异(p > 0.06)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To compare image quality and radiation doses for chest DECT acquired with dual-source and rapid-kV switching techniques.

MATERIALS AND METHODS: Our institutional Review Board approved retrospective study included 97 patients (54 men, 43 women; 63 ± 14 years) who underwent contrast-enhanced chest DECT with both single source, rapid kV-switching (SS-DECT) and dual source (DS-DECT) techniques per standard of care departmental protocols. Reconstructed images from both scanners had identical section thickness and section interval for virtual monoenergetic and material decomposition iodine (MDI) images. Two thoracic radiologists independently evaluated all DECT for findings, quality of images, perfusion defects (MDI), and presence of artifacts. Radiation dose descriptor, size-specific dose estimates (SSDE), was recorded. Data were analyzed with Wilcoxon Signed Rank and Cohen's Kappa tests.

RESULTS: There were no significant differences in patient weight or SSDE for the two DECT techniques (p > 0.06). Both radiologists reported no difference in lesion and artifact evaluation on the virtual monoenergetic images from either technique (p > 0.05). However, SS-DECT (in 63-71/97 patients) had substantial artifactual heterogeneity in pulmonary perfusion on MDI images compared to none on DS-DECT (p < 0.001).

CONCLUSION: Despite identical patients and associated radiation doses, there were substantial differences in material decomposition iodine images generated from SS-DECT and DS-DECT techniques. Pulmonary heterogeneity on MDI images from SS-DECT leads to artifactual areas of low perfusion and can confound interpretation of true pulmonary perfusion.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Singh R, Sharma A, McDermott S, et al. Comparison of image quality and radiation doses between rapid kV-switching and dual-source DECT techniques in the chest. Eur J Radiol. 2019. PMID: 31442929. doi:10.1016/j.ejrad.2019.08.008