使用骨抑制技术和低管电压降低胸部X线摄影剂量并改善图像质量:一项体模研究

Dose reduction and image quality improvement of chest radiography by using bone-suppression technique and low tube voltage: a phantom study.
PubMed 文献 · Eur Radiol · 2019-08-05

本研究使用胸部体模和模拟肺结节,在四种入射表面剂量条件下采集264幅胸片并生成骨抑制图像,分析结节对比度和对比噪声比。结果显示,在非肋骨重叠区域,骨抑制技术显著提高对比度和CNR(p<0.01);骨抑制图像中70 kVp三种条件的数值等于或显著高于140 kVp条件;70 kVp下约0.3 mGy与0.1 mGy之间无明显下降。结论认为,在不过度增加模糊伪影和图像噪声的前提下,使用最短曝光时间和最低管电压可改善骨抑制图像质量并降低患者剂量。

要点

  • 在非肋骨重叠区域,骨抑制技术使结节对比度和CNR显著提高(p<0.01)。
  • 骨抑制图像中70 kVp三种条件的对比度和CNR等于或显著高于140 kVp条件。
  • 70 kVp下入射表面剂量约0.3 mGy与0.1 mGy之间对比度和CNR无明显下降。

核心结论

  • 在不过度增加模糊伪影和图像噪声的前提下,使用最短曝光时间和最低管电压可改善骨抑制图像质量并降低患者剂量。

研究设计与样本

  • 体模研究,使用拟人胸部体模和直径12 mm、CT值约+100 HU的球形模拟结节。
  • 肺野分为7个区域,结节置于共66个位置,共采集264幅胸片。
  • 四种入射表面剂量条件:140 kVp下约0.3 mGy、70 kVp下约0.3 mGy、70 kVp下约0.2 mGy、70 kVp下约0.1 mGy。
  • 图像经处理生成骨抑制图像,采用双侧Wilcoxon符号秩检验分析结节对比度和CNR差异。

诊断标准与关键数值

  • 在非肋骨重叠区域,骨抑制技术使结节对比度和CNR显著增加(p<0.01)。
  • 骨抑制图像中70 kVp三种条件的对比度和CNR等于或显著高于140 kVp条件。
  • 70 kVp下入射表面剂量约0.3 mGy与0.1 mGy之间对比度和CNR无明显下降。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To clarify the relationship between entrance surface dose (ESD) and physical image quality of original and bone-suppressed chest radiographs acquired using high and low tube voltages.

METHODS: An anthropomorphic chest phantom and a 12-mm diameter spherical simulated nodule with a CT value of approximately + 100 HU were used. The lung field in the chest radiograph was divided into seven areas, and the nodule was set in a total of 66 positions. A total of 264 chest radiographs were acquired using four ESD conditions: approximately 0.3 mGy at 140 and 70 kVp and approximately 0.2 and 0.1 mGy at 70 kVp. The radiographs were processed to produce bone-suppressed images. Differences in contrast and contrast-to-noise ratio (CNR) values of the nodule between each condition and between the original and bone-suppressed images were analyzed by a two-sided Wilcoxon signed-rank test.

RESULTS: In the areas not overlapping with the ribs, both contrast and CNR values were significantly increased with the bone-suppression technique (p < 0.01). In the bone-suppressed images, these values of the three conditions at 70 kVp were equal to or significantly higher than those of the condition at 140 kVp. There was no apparent decrease in these values between the ESD of approximately 0.3 and 0.1 mGy at 70 kVp.

CONCLUSION: By using the shortest exposure time and the lowest tube voltage possible not to increase in blurring artifact and image noise, it is possible to improve the image quality of bone-suppressed images and reduce the patient dose.

KEY POINTS: • The effectiveness of bone-suppression techniques differs in areas of lung field. • Image quality of bone-suppressed chest radiographs is improved by lower tube voltage. • Applying lower tube voltage to bone-suppressed chest radiographs leads to dose reduction.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Takagi S, Yaegashi T, Ishikawa M, et al. Dose reduction and image quality improvement of chest radiography by using bone-suppression technique and low tube voltage: a phantom study. Eur Radiol. 2019. PMID: 31385049. doi:10.1007/s00330-019-06375-6

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。