超音速剪切波弹性成像在大鼠中诱发的肺毛细血管出血

Pulmonary Capillary Hemorrhage Induced by Super Sonic Shear Wave Elastography in Rats.
PubMed 文献 · Ultrasound Med Biol · 2019-08-12

本研究在大鼠中观察了SuperSonic Imagine剪切波弹性成像系统(Aixplorer)诊断超声的肺毛细血管出血(PCH)生物效应。弹性成像以1 Hz重复,包含间隔较大的B模式脉冲、超音速推脉冲(SSP)和剪切波成像(SWI)脉冲。结果表明B模式仅在最0 dB输出时产生显著PCH,而SSP联合SWI脉冲在300秒、30秒甚至单帧曝光时均产生显著PCH,300秒曝光的峰值稀疏压振幅阈值为1.5 MPa(MIIS=0.67),30秒为1.7 MPa(MIIS=0.76),麻醉中去除赛拉嗪后阈值升至2.1 MPa(MIIS=0.94)。

要点

  • SSP联合SWI脉冲比B模式更易诱发肺毛细血管出血,单帧曝光即可产生显著PCH。
  • 300秒曝光PCH阈值1.5 MPa(MIIS=0.67),30秒曝光阈值1.7 MPa(MIIS=0.76)。
  • 麻醉中去除赛拉嗪后PCH阈值升高至2.1 MPa(MIIS=0.94)。

核心结论

  • SSP脉冲联合SWI脉冲比B模式更有效地诱发大鼠肺毛细血管出血,但其阈值与其他系统其他诊断超声模式的既往结果相当。

研究设计与样本

  • 动物实验研究,使用SuperSonic Imagine剪切波弹性成像系统(Aixplorer,Supersonic Imagine,Aix-en-Provence,法国)对大鼠进行右体侧温水浴成像,弹性成像以1 Hz重复,包含间隔较大的B模式图像脉冲、超音速推(SSP)脉冲和剪切波成像(SWI)脉冲;大鼠用氯胺酮和赛拉嗪或仅用氯胺酮麻醉,分为单帧、30秒和300秒曝光组,另设仅低功率瞄准扫描的假组;曝光后5分钟取肺评估PCH面积和体积。

诊断标准与关键数值

  • 300秒曝光时,无论是否加入B模式脉冲,峰值稀疏压振幅PCH阈值均为1.5 MPa(原位机械指数MIIS=0.67)。
  • 30秒曝光时阈值略升高至1.7 MPa(MIIS=0.76)。
  • 麻醉中去除赛拉嗪后阈值升高至2.1 MPa(MIIS=0.94)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

The occurrence of the pulmonary capillary hemorrhage (PCH) bioeffect of diagnostic ultrasound in rats was investigated for a SuperSonic Imagine shear wave elastography system (Aixplorer, Supersonic Imagine, Aix-en-Provence, France). The elastography imaging repeated at 1 Hz and consisted of widely spaced B-mode image pulses, supersonic push (SSP) pulses and shear wave imaging (SWI) pulses. Groups of rats anesthetized with ketamine and xylazine, or with ketamine only, were imaged on their right side in a warm water bath for one frame, 30 s and 300 s. The image focus and region of interest were adjusted to give exposure only with the background B-mode imaging, or primarily with the SSP and SWI pulses. A sham group had only low power aiming scans. The lungs were removed 5 min after exposure and evaluated for PCH area and volume. The B mode was notably ineffective and produced significant PCH only at the maximum 0 dB output. The SSP pulses together with the SWI pulses produced significant PCH for 300 s, 30 s and even single image exposures. Peak rarefactional pressure amplitude PCH thresholds for 300 s exposure were the same with or without the B-mode pulses at 1.5 MPa (in situ mechanical index, MIIS = 0.67). A 30 s duration resulted in a slightly increased threshold of 1.7 MPa (MIIS = 0.76). The omission of xylazine from the anesthetic, which reduces the sensitivity of rat lung to PCH occurrence, resulted in an increased threshold of 2.1 MPa (MIIS = 0.94). The unique SSP pulses were much more effective than the B mode, but thresholds were comparable to previous results with other diagnostic ultrasound modes on other systems.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。