CT金属牙科伪影减少技术(Smart MAR):对疑似头颈部病变及口腔植入物患者图像质量和诊断信心的改善
研究纳入100例有不可移除口腔植入物或牙科填充物、疑似口咽癌或脓肿的患者,对单源多层CT原始数据分别行ASIR V单独重建与附加Smart MAR重建。附加Smart MAR可显著改善植入物邻近及较远区域的衰减和噪声、SNR和CNR,并改善肿瘤/脓肿显示;36例中8例(22%)肿瘤仅在Smart MAR序列中检出。
要点
- 附加Smart MAR重建较单独ASIR V显著改善植入物邻近及较远区域的客观图像质量(衰减、噪声均p<0.001)。
- Smart MAR改善肿瘤/脓肿可视化及代表性口咽组织显示(p<0.001),36例中8例(22%)肿瘤仅在Smart MAR序列检出。
- 疑似头颈部病变伴口腔植入物患者行CT时,附加Smart MAR后处理有助于提高诊断信心和病灶检出。
核心结论
- 附加使用Smart MAR后处理工具似乎可显著改善头颈部CT的主客观图像质量、诊断信心和病灶检出。
- 22%的病例中肿瘤仅在Smart MAR重建图像中检出。
研究设计与样本
- 研究连续纳入100例有不可移除口腔植入物或牙科填充物、疑似口咽癌或脓肿的患者。
- 对单源多层CT扫描仪的CT原始数据分别行ASIR V单独重建与附加Smart MAR重建。
- Smart MAR为三阶段、基于投影的后处理算法;图像质量由5个感兴趣区定量比较及两名独立评估者定性比较。
关键影像表现/征象
- 附加Smart MAR重建较单独ASIR V重建似乎显著改善植入物邻近及较远区域的衰减和噪声(均p<0.001)。
- Smart MAR改善肿瘤/脓肿(100例中检出36例,36%)及代表性口咽组织的可视化(p<0.001)。
- 36例肿瘤患者中8例(22%)肿瘤仅在Smart MAR序列中检出。
诊断标准与关键数值
- 与单独ASIR V相比,附加Smart MAR后SNR改善(p=0.001),CNR改善(p=0.001)。
- 定量分析采用5个感兴趣区,定性分析由两名独立评估者完成。
- 平均总DLP为506.8 mGy*cm;平均CTDIvol为5.5 mGy。
临床应用与阅片要点
- 对伴有口腔植入物或牙科填充物、疑似头颈部病变的患者,在ASIR V重建基础上附加Smart MAR后处理似乎可提高图像质量和诊断信心。
- 阅片时应注意肿瘤可能仅在Smart MAR重建序列中显示,需结合该序列评估以免漏诊。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: We determined whether the Smart MAR metal artifact reduction tool - a three-stage, projection-based, post processing algorithm - improves subjective and objective image quality and diagnostic confidence in patients with dental artifacts and suspected head and neck pathology compared to standard adaptive statistical iterative reconstructions (ASIR V) alone.
METHOD: The study included 100 consecutive patients with nonremovable oral implants or dental fillings and suspected oropharyngeal cancer or abscess. CT raw data of a single-source multislice CT scanner were postprocessed using ASIR V alone and with additional Smart MAR reconstruction. Image quality of baseline ASIR V and Smart MAR-based reconstruction series was compared both quantitatively (5 regions of interest, ROIs) and qualitatively (two independent raters).
RESULTS: Additional Smart MAR reconstruction significantly seems to improve both attenuation and noise adjacent to implants and in more distant areas (all p < 0.001) compared to standard ASIR V reconstructions alone. Signal-to-noise ratio (SNR; p = 0.001) and contrast-to-noise ratio were improved significantly (CNR; p = 0.001). Smart MAR improved visualization of tumor/abscess (detected in 36 of 100 patients, 36%) and representative oropharyngeal tissue (p < 0.001). In 8 of 36 patients (22%), tumor was only detected in Smart MAR series. Mean total DLP was 506.8mGy*cm; average CTDIvol was 5.5 mGy.
CONCLUSIONS: The supplementary use of the Smart MAR post-processing tool seems to significantly improve both subjective and objective image quality as well as diagnostic confidence and lesion detection in CT of the head and neck. In 22% of cases, the tumor was detected only in Smart MAR reconstructed images.
引用
- Feldhaus F, Böning G, Jonczyk M, et al. Metallic dental artifact reduction in computed tomography (Smart MAR): Improvement of image quality and diagnostic confidence in patients with suspected head and neck pathology and oral implants. Eur J Radiol. 2019. PMID: 31439235. doi:10.1016/j.ejrad.2019.07.015
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。