容积高分辨率重T2加权成像与时间分辨对比增强MRA在脊髓血管畸形评估中的对比分析
前瞻性纳入28例脊髓血管畸形患者,由2名神经放射科医师盲法分别分析容积高分辨率重T2加权成像与时间分辨对比增强MRA。两种序列检测脊髓血管畸形的敏感度均为100%、准确率均为93.5%,联合评估可提高所有类型脊髓血管畸形的总体诊断准确率及供血动脉特征描述的准确性。
要点
- 容积高分辨率重T2加权成像与时间分辨对比增强MRA检测脊髓血管畸形的敏感度均为100%、准确率均为93.5%。
- 脊髓动静脉畸形容积高分辨率重T2加权成像敏感度100% versus 时间分辨对比增强MRI 90%,准确率96.7% versus 93.5%。
- 联合评估可提高所有类型脊髓血管畸形的总体诊断准确率,并改善供血动脉特征描述。
核心结论
- 联合容积高分辨率重T2加权成像与时间分辨对比增强MRI可提高脊髓血管畸形诊断、分类及供血动脉特征描述的敏感度和准确率。
研究设计与样本
- 前瞻性纳入28例脊髓血管畸形患者(脊髓硬脊膜动静脉瘘、脊髓动静脉畸形、髓周动静脉瘘),由2名对最终诊断盲法的神经放射科医师分别独立分析MRI图像。
诊断标准与关键数值
- 检测脊髓血管畸形:两种序列敏感度均为100%、准确率均为93.5%。
- 脊髓动静脉畸形:容积高分辨率重T2加权成像敏感度100% versus 时间分辨对比增强MRI 90%,准确率96.7% versus 93.5%。
- 髓周动静脉瘘:时间分辨对比增强MRI敏感度80% versus 容积T2序列60%,准确率96.7% versus 93.5%。
- 脊髓硬脊膜动静脉瘘:两种序列敏感度均为100%、准确率均为87%。
- 脊髓硬脊膜动静脉瘘供血动脉识别:容积高分辨率重T2加权成像对两名观察者均优于或等于时间分辨对比增强MRI(90.9%和72.7% versus 72.7%),联合序列时提高至90.9%。
临床应用与阅片要点
- 作者建议联合容积高分辨率重T2加权成像与时间分辨对比增强MRI用于脊髓血管畸形的诊断、分类及供血动脉特征描述。
- 两种序列对脊髓硬脊膜动静脉瘘检测敏感度相同,但容积高分辨率重T2加权成像在供血动脉识别上优于或等于时间分辨对比增强MRI。
- 时间分辨对比增强MRI对髓周动静脉瘘及脊髓动静脉畸形供血动脉识别优于容积高分辨率重T2加权成像。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Volumetric high-resolution heavily T2-weighted imaging or time-resolved contrast-enhanced MRA is used in the detection and characterization of spinal vascular malformations, though inherent trade-offs can affect their overall sensitivity and accuracy. We compared the efficacy of volumetric high-resolution heavily T2-weighted and time-resolved contrast-enhanced images in spinal vascular malformation diagnosis and feeder characterization and assessed whether a combined evaluation improved the overall accuracy of diagnosis.
MATERIALS AND METHODS: Twenty-eight patients with spinal vascular malformations (spinal dural arteriovenous fistula, spinal cord arteriovenous malformation, and perimedullary arteriovenous fistula) were prospectively enrolled. MR images were separately analyzed by 2 neuroradiologists blinded to the final diagnosis.
RESULTS: Both sequences demonstrated 100% sensitivity and 93.5% accuracy for the detection of spinal vascular malformations. Volumetric high-resolution heavily T2-weighted imaging was superior to time-resolved contrast-enhanced MR imaging for identification of spinal cord arteriovenous malformations (100% versus 90% sensitivity and 96.7% versus 93.5% accuracy), however, for the diagnosis of perimedullary arteriovenous fistula, time resolved contrast enhanced MRI was found to perform better than the volumetric T2 sequence (80% versus 60% sensitivity and 96.7% versus 93.5% accuracy). Both sequences showed equal sensitivity (100%) and accuracy (87%) for spinal dural arteriovenous fistulas. Combined evaluation improved the overall accuracy across all types of spinal vascular malformation. Volumetric high-resolution heavily T2-weighted imaging was superior or equal to time-resolved contrast-enhanced MR imaging for feeder identification of spinal dural arteriovenous fistulas for both observers (90.9% and 72.7% versus 72.7%), which improved to 90.9% when the sequences were combined. Time-resolved contrast-enhanced MR imaging performed better for major and total feeder identification of spinal cord arteriovenous malformation (80% versus 60%) and perimedullary arteriovenous fistula (80% versus 60%-80%).
CONCLUSIONS: Combined volumetric high-resolution heavily T2-weighted imaging and time-resolved contrast-enhanced MR imaging can improve the sensitivity and accuracy of spinal vascular malformation diagnosis, classification, and feeder characterization.