成人肿瘤CT中患者特异性器官剂量与有效剂量的估算
本研究随机收集2年内接受CT检查的1200例成人肿瘤患者数据,利用患者体型与扫描技术参数结合计算人体模体估算器官剂量与有效剂量。脑扫描眼晶状体平均剂量59.9 mGy,胸部扫描甲状腺10.1 mGy,腹盆部扫描肝脏17.5 mGy,体部扫描肝脏19.0 mGy;不同体型患者剂量估算相差可达2倍。
要点
- 脑扫描眼晶状体平均剂量59.9 mGy,为报告中需关注的器官剂量。
- 胸部扫描甲状腺平均剂量10.1 mGy,腹盆部及体部扫描肝脏剂量分别17.5与19.0 mGy。
- 不同体型患者剂量估算相差可达2倍,提示报告剂量时需结合患者体型。
核心结论
- 可利用患者体型与扫描数据快速估算检查和患者特异性的器官剂量与有效剂量,与通用剂量估算方法相比可相差2倍。
研究设计与样本
- 随机收集2年内接受CT检查的1200例成人肿瘤患者数据,以患者体格特征和CT技术参数为输入,结合按性别和有效直径匹配的计算人体模体估算器官剂量与有效剂量。
诊断标准与关键数值
- 脑扫描眼晶状体平均剂量59.9 mGy。
- 胸部扫描甲状腺平均剂量10.1 mGy。
- 腹盆部扫描肝脏平均剂量17.5 mGy。
- 体部扫描肝脏平均剂量19.0 mGy。
- 不同体型患者间剂量估算相差可达2倍。
临床应用与阅片要点
- 可利用患者体型与扫描参数估算检查和患者特异性器官剂量与有效剂量。
- 不同体型患者剂量估算可相差2倍,报告剂量时需考虑体型差异。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE. Patient-specific organ and effective dose provides essential information for CT protocol optimization. However, such information is not readily available in the scan records. The purpose of this study was to develop a method to obtain accurate examination- and patient-specific organ and effective dose estimates by use of available scan data and patient body size information for a large cohort of patients. MATERIALS AND METHODS. The data were randomly collected for 1200 patients who underwent CT in a 2-year period. Physical characteristics of the patients and CT technique were processed as inputs for the dose estimator. Organ and effective doses were estimated by use of the inputs and computational human phantoms matched to patients on the basis of sex and effective diameter. Size-based ratios were applied to correct for patient-phantom body size differences. RESULTS. Patients received a mean of 59.9 mGy to the lens of the eye per brain scan, 10.1 mGy to the thyroid per chest scan, 17.5 mGy to the liver per abdomen and pelvis scan, and 19.0 mGy to the liver per body scan. A factor of 2 difference in dose estimates was observed between patients of various habitus. CONCLUSION. Examination- and patient-specific organ and effective doses were estimated for 1200 adult oncology patients undergoing CT. The dose conversion factors calculated facilitate rapid organ and effective dose estimation in clinics. Compared with nonspecific dose estimation methods, patient dose estimations with data specific to the patient and examination can differ by a factor of 2.
引用
- Gao Y, Mahmood U, Liu T, et al. Patient-Specific Organ and Effective Dose Estimates in Adult Oncologic CT. AJR Am J Roentgenol. 2019. PMID: 31414882. doi:10.2214/AJR.19.21197