深部脑刺激期间的脑功能磁共振成像安全性与伪影:102例患者经验
该前瞻性研究纳入102例接受活动性深部脑刺激(DBS)的患者,在2017年6月至2018年10月间进行1.5 T或3 T磁共振成像(T1加权成像和梯度回波-回波平面成像),追踪短期和长期不良事件,并量化DBS硬件相关伪影。结果显示无MRI相关短期或长期不良事件或急性改变,伪影主要位于电极触点附近及顶额叶皮质延伸导线冗余处,3 T线圈伪影较1.5 T更大(分别遮蔽2.1%±0.7和1.4%±0.7颅内体积,P<.001)。作者结论认为,经过事先局部安全测试,DBS患者可安全接受1.5 T和3 T MRI,硬件相关伪影仅影响小部分脑区。
要点
- DBS患者经事先局部安全测试后,可安全进行1.5 T和3 T MRI,未观察到短期或长期不良事件。
- 功能MRI(GRE-EPI)上DBS伪影最显著位于电极触点附近和顶额叶皮质皮下延伸导线冗余处。
- 3 T线圈伪影遮蔽颅内体积2.1%±0.7,大于1.5 T的1.4%±0.7(P<.001),主要遮蔽浅表顶额叶皮质和电极触点邻近深部结构。
核心结论
- 经过事先局部安全测试,接受深部脑刺激的患者可以安全地进行1.5 T和3 T MRI。
- 深部脑刺激硬件相关伪影仅影响小部分脑区。
研究设计与样本
- 前瞻性研究,纳入102例接受活动性DBS的参与者(平均年龄60岁±11,65例男性),在2017年6月至2018年10月间进行1.5 T或3 T MRI(T1加权成像和梯度回波-回波平面成像)。
- 追踪短期和长期不良事件,量化GRE-EPI(功能MRI)上颅骨线圈导线和电极触点处的DBS硬件相关伪影,并将分割伪影转换至标准空间以定义信号丢失影响的脑区。
- 采用两样本t检验评估1.5 T与3 T MRI之间伪影大小的差异。
关键影像表现/征象
- DBS伪影在电极触点附近以及皮下延伸导线冗余处所在的顶额叶皮质区域最为显著。
- 电极触点伪影平均直径为9.3 mm±1.6,线圈伪影遮蔽脑组织的1.9%±0.8。
- 线圈伪影在3 T较1.5 T更大,分别遮蔽颅内体积的2.1%±0.7和1.4%±0.7(P<.001)。
- 浅表顶额叶皮质和电极触点邻近深部结构最常被遮蔽。
诊断标准与关键数值
- 电极触点伪影平均直径为9.3 mm±1.6。
- 线圈伪影遮蔽脑组织的1.9%±0.8。
- 3 T线圈伪影遮蔽颅内体积2.1%±0.7,1.5 T为1.4%±0.7,3 T较1.5 T更大(P<.001)。
临床应用与阅片要点
- 对于接受活动性DBS的患者,需在事先局部安全测试的前提下进行1.5 T或3 T MRI。
- 阅片时应注意电极触点附近及顶额叶皮质皮下延伸导线冗余处的伪影,这些区域信号丢失可能影响邻近脑组织的评估。
- 3 T时线圈伪影较1.5 T更大,遮蔽更多颅内体积,选择场强时需考虑伪影范围。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BackgroundWith growing numbers of patients receiving deep brain stimulation (DBS), radiologists are encountering these neuromodulation devices at an increasing rate. Current MRI safety guidelines, however, limit MRI access in these patients.PurposeTo describe an MRI (1.5 T and 3 T) experience and safety profile in a large cohort of participants with active DBS systems and characterize the hardware-related artifacts on images from functional MRI.Materials and MethodsIn this prospective study, study participants receiving active DBS underwent 1.5- or 3-T MRI (T1-weighted imaging and gradient-recalled echo [GRE]-echo-planar imaging [EPI]) between June 2017 and October 2018. Short- and long-term adverse events were tracked. The authors quantified DBS hardware-related artifacts on images from GRE-EPI (functional MRI) at the cranial coil wire and electrode contacts. Segmented artifacts were then transformed into standard space to define the brain areas affected by signal loss. Two-sample t tests were used to assess the difference in artifact size between 1.5- and 3-T MRI.ResultsA total of 102 participants (mean age ± standard deviation, 60 years ± 11; 65 men) were evaluated. No MRI-related short- and long-term adverse events or acute changes were observed. DBS artifacts were most prominent near the electrode contacts and over the frontoparietal cortical area where the redundancy of the extension wire is placed subcutaneously. The mean electrode contact artifact diameter was 9.3 mm ± 1.6, and 1.9% ± 0.8 of the brain was obscured by the coil artifact. The coil artifacts were larger at 3 T than at 1.5 T, obscuring 2.1% ± 0.7 and 1.4% ± 0.7 of intracranial volume, respectively (P < .001). The superficial frontoparietal cortex and deep structures neighboring the electrode contacts were most commonly obscured.ConclusionWith a priori local safety testing, patients receiving deep brain stimulation may safely undergo 1.5- and 3-T MRI. Deep brain stimulation hardware-related artifacts only affect a small proportion of the brain.© RSNA, 2019Online supplemental material is available for this article.See also the editorial by Martin in this issue.
引用
- Boutet A, Rashid T, Hancu I, et al. Functional MRI Safety and Artifacts during Deep Brain Stimulation: Experience in 102 Patients. Radiology. 2019. PMID: 31385756. doi:10.1148/radiol.2019190546
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。