脑微出血的MRI检测:大小至关重要
本研究为影像-组织病理学对照研究,纳入42例连续尸检病例,在临床3T系统上采集死后SWI图像并与对应组织病理切片对照。死后MRI检出23个真阳性脑微出血,组织病理学另检出68个假阴性脑微出血;真阳性病灶平均大小3.6±7.1 mm³,假阴性病灶平均0.3±1.2 mm³,显著更小(p<0.05)。
要点
- MRI真阳性脑微出血的组织病理平均大小为3.6±7.1 mm³。
- MRI假阴性脑微出血的组织病理平均大小为0.3±1.2 mm³,显著更小(p<0.05)。
- 真阳性脑微出血平均大小约为假阴性的10倍。
核心结论
- 脑微出血大小至关重要,真阳性MRI脑微出血平均大小约为假阴性的10倍;除大小外,铁含量、Fe2+/Fe3+比值、空间构型和位置等因素也可能影响检出性,尚待未来研究阐明。
研究设计与样本
- 影像-组织病理学对照研究,经当地审查委员会批准,纳入2012年1月12日至2012年12月10日间42例连续尸检病例(死亡时平均年龄80.7±10.0岁,女性23例、男性19例),在临床3T系统上以类似临床常规的参数采集死后SWI图像,并与对应组织病理切片对照。
诊断标准与关键数值
- 死后MRI检出23个真阳性脑微出血,组织病理学另检出68个假阴性脑微出血。
- 真阳性MRI脑微出血在组织病理上的平均大小为3.6±7.1 mm³。
- 假阴性MRI脑微出血在组织病理上的平均大小为0.3±1.2 mm³,显著小于真阳性(p<0.05)。
局限性
- 原文指出除大小外,铁含量、Fe2+/Fe3+比值、空间构型和位置等其他因素也会影响脑微出血的可检出性,尚待未来研究阐明。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: Cerebral microbleeds (CMB) play an important role as an imaging biomarker notably in vascular and neurodegenerative diseases. Current clinical brain MRI underestimates the number of CMB with respect to histopathology. It is expected that small CMBs are more likely to be false-negatives, yet this has not been demonstrated and the average size of false-negative and true-positive CMBs have not been established.
METHODS: The radiologic-histopathologic correlation study was approved by the local review board and included 42 consecutive cases (mean age at death, 80.7 ± 10.0 years; 23 females and 19 men) between 12 January 2012 and 10 December 2012 having undergone brain autopsy. Postmortem SWI (susceptibility-weighted imaging) images were acquired on a clinical 3T system using parameters similar to clinical routine. The detection of CMB on postmortem MRI was compared with corresponding histopathological slices.
RESULTS: Postmortem MRI detected 23 true-positive CMB. Histopathology additionally detected 68 CMBs (false-negative MRI CMBs). The average size true-positive MRI CMBs had on histopathology was 3.6 ± 7.1 mm3. The average size false-negative MRI CMBs was significantly smaller (p < 0.05), measuring 0.3 ± 1.2 mm3 on histopathology.
CONCLUSION: Size matters. As expected, the average size of true-positive MRI CMB was around 10 times larger as compared with false-negative MRI CMB. Evidently, in addition to size, other factors will influence the detectability of CMB, including iron content, ratio of Fe2+/Fe3+, spatial configuration, and location, yet this remains to be elucidated in future studies.