改良钡餐吞咽检查中脉冲率与辐射剂量的权衡:现实情况如何
该回顾性研究纳入6个月内住院患者连续改良钡餐吞咽(MBS)检查,比较实施MBS损伤量表(MBSImP)前后两个队列(低脉冲率/正常剂量 vs 高脉冲率/低剂量)的患者辐射剂量与透视时间。体模测试显示两种模式剂量无显著差异;标准化后平均辐射剂量从5.86 mGy降至4.72 mGy(p<0.0001),透视时间无显著变化(p=0.62),剂量率从4.35降至3.55 mGy/s(p<0.0001)。结论认为降低剂量模式并提高透视帧率可降低患者辐射剂量且不增加透视时间。
要点
- 实施MBSImP后采用高帧率/低剂量模式,平均辐射剂量由5.86 mGy降至4.72 mGy(p<0.0001)。
- 标准化前后平均透视时间无显著差异(83.8秒 vs 82.3秒,p=0.62)。
- 体模测试显示低帧率/正常剂量与高帧率/低剂量模式剂量无显著差异。
核心结论
- 降低剂量模式并提高透视帧率可降低患者辐射剂量,且不增加透视时间。
研究设计与样本
- 回顾性研究,纳入6个月内住院患者连续MBS检查,分实施MBSImP前(低帧率、正常剂量)449例与实施后(高帧率、低剂量)378例两个队列。
- 实施前在多个体模上进行脉冲率与剂量测试。
诊断标准与关键数值
- 体模剂量测试显示低帧率/正常剂量与高帧率/低剂量模式之间剂量无显著差异。
- 实施前平均辐射剂量5.86(±4.35)mGy,实施后4.72(±3.77)mGy(p<0.0001)。
- 实施前平均透视时间83.8(±44.4)秒,实施后82.3(±39.8)秒(p=0.62)。
- 实施前剂量率4.35(±2.42)mGy/s,实施后3.55(±2.41)mGy/s(p<0.0001)。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To evaluate the effect of modification of dose mode and frame rate on patient radiation dose during modified barium swallow (MBS) examinations.
MATERIALS AND METHODS: A retrospective review was undertaken of consecutive MBS examinations performed over 6 months in the inpatient setting. Patients were divided into two cohorts: pre-implementation of the MBS Impairment Profile (MBSImP; low rate, normal dose) and post-implementation (high rate, low dose). Prior to implementation, pulse rate and dose testing were performed on multiple phantoms.
RESULTS: Four hundred and forty-nine patients were included in the pre-implementation cohort and 378 in the post-implementation cohort. Phantom dose testing demonstrated no significant difference in dose on either phantom between low rate/normal dose and high rate/low dose modes. Prior to MBS standardisation, the mean radiation dose was 5.86 (±4.35) mGy. Following standardisation, the mean radiation dose was 4.72 (±3.77) mGy (p<0.0001). The mean fluoroscopy time for MBS prior to standardisation was 83.8 (±44.4) seconds and the mean fluoroscopy time for MBS after standardisation was 82.3 (±39.8) seconds (p=0.62). The dose rate for MBS prior to standardisation was 4.35 (±2.42) and the dose rate for MBS after standardisation was 3.55 (±2.41) mGy/s (p<0.0001).
CONCLUSION: Adjustments made to lower the dose mode and the increase in fluoroscopy frame rate decreased the patient radiation dose and did not increase fluoroscopy time.
引用
- Galgano SJ, Gauntt D, Boyd MR, et al. Trade-off between pulse rate and radiation dose during modified barium swallow examination: what is the reality? Clin Radiol. 2019. PMID: 31256907. doi:10.1016/j.crad.2019.05.030
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。