自由呼吸快速小角度激发静息间期层选择性磁共振血管成像技术提升盆腔和腹部血管狭窄检出率的技术开发
该前瞻性研究纳入27例外周动脉疾病患者,在3T下比较标准屏气bSSFP QISS与自由呼吸径向FLASH及笛卡尔FLASH QISS的无对比剂MRA图像质量和诊断准确性,以CT血管造影为参考。结果显示径向FLASH QISS图像质量最佳(P<0.0001)且读者间一致性最高(κ=0.721),FLASH QISS对>50%狭窄的敏感度高于bSSFP QISS,并持续减少偏共振伪影。
要点
- 自由呼吸径向FLASH QISS图像质量最佳(P<0.0001),读者间一致性最高(κ=0.721)。
- FLASH QISS检测>50%狭窄的敏感度(径向84.0%、笛卡尔80.0%)高于bSSFP QISS(76.0%)。
- FLASH QISS较bSSFP QISS持续减少偏共振伪影。
核心结论
- 自由呼吸FLASH QISS MRA技术为腹部和盆腔血管狭窄检测提供改善的图像质量和敏感度、高特异度并减少偏共振伪影。
研究设计与样本
- 机构审查委员会批准、符合健康保险便携与责任法案的前瞻性研究,2018年4月至12月纳入27例外周动脉疾病患者(69±10岁,17名男性)。
- 患者接受3T无对比剂MRA,使用标准bSSFP QISS和原型自由呼吸径向FLASH及笛卡尔FLASH QISS。
- 其中22例患者还接受CT血管造影作为参考标准,每例评估腹部、盆腔和大腿上段共9个动脉节段。
关键影像表现/征象
- FLASH QISS较bSSFP QISS持续减少偏共振伪影。
- 径向FLASH QISS图像质量最佳(P<0.0001),读者间一致性最高(κ=0.721)。
诊断标准与关键数值
- bSSFP QISS检测>50%狭窄的敏感度为76.0%、特异度为97.6%。
- 径向FLASH QISS检测>50%狭窄的敏感度为84.0%、特异度为94.0%。
- 笛卡尔FLASH QISS检测>50%狭窄的敏感度为80.0%、特异度为92.8%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Balanced steady-state free precession-based quiescent-interval slice-selective (bSSFP QISS) magnetic resonance angiography (MRA) is accurate for the noncontrast evaluation of peripheral arterial disease (PAD); however, drawbacks include the need for breath-holding when imaging the abdomen and pelvis, and sensitivity to off-resonance artifacts. The purpose of this study was to evaluate the image quality and diagnostic accuracy in the pelvis and abdomen of free-breathing fast low-angle shot-based QISS (FLASH QISS) techniques in comparison to bSSFP QISS in patients with PAD, using computed tomographic angiography as the reference.
MATERIALS AND METHODS: Twenty-seven patients (69 ± 10 years, 17 men) with PAD were enrolled in this institutional review board-approved, Health Insurance Portability and Accountability Act-compliant prospective study between April and December 2018. Patients underwent noncontrast MRA using standard bSSFP QISS and prototype free-breathing radial-FLASH and Cartesian-FLASH QISS at 3 T. A subset of patients (n = 22) also underwent computed tomographic angiography as the reference standard. Nine arterial segments per patient were evaluated spanning the abdomen, pelvis, and upper thigh regions. Objective (signal intensity ratio and relative standard deviation) and subjective image quality (4-point scale) and stenosis (>50%) were evaluated by 2 readers and compared using one-way analysis of variance, Wilcoxon, and McNemar tests, respectively.
RESULTS: A total of 179 vascular segments were available for analysis by all QISS techniques. No significant difference was observed among bSSFP, radial-FLASH, and Cartesian-FLASH QISS techniques in signal intensity ratio (P = 0.428) and relative standard deviation (P = 0.220). Radial-FLASH QISS demonstrated the best image quality (P < 0.0001) and the highest interreader agreement (κ = 0.721). The sensitivity values of bSSFP, radial-FLASH, and Cartesian-FLASH QISS for the detection of greater than 50% stenosis were 76.0%, 84.0%, and 80.0%, respectively, whereas specificity values were 97.6%, 94.0%, and 92.8%, respectively. Moreover, FLASH QISS consistently reduced off-resonance artifacts compared with bSSFP QISS.
CONCLUSIONS: Free-breathing FLASH QISS MRA techniques provide improved image quality and sensitivity, high specificity, and reduced off-resonance artifacts for vascular stenosis detection in the abdomen and pelvis.
引用
- Varga-Szemes A, Aherne EA, Schoepf UJ, et al. Free-Breathing Fast Low-Angle Shot Quiescent-Interval Slice-Selective Magnetic Resonance Angiography for Improved Detection of Vascular Stenoses in the Pelvis and Abdomen: Technical Development. Invest Radiol. 2019. PMID: 31299678. doi:10.1097/RLI.0000000000000592