80 kVp 基于模型的迭代重建与 70 kVp 自适应统计迭代重建-V 在下肢 CT 静脉造影中的图像质量与辐射剂量比较

Image Quality and Radiation Dose in CT Venography Using Model-Based Iterative Reconstruction at 80 kVp versus Adaptive Statistical Iterative Reconstruction-V at 70 kVp.
PubMed 文献 · Korean J Radiol · 2019-01-01

B 组血管强化、SNR、CNR 及多数主观图像质量指标优于 A 组,重建时间显著缩短(1 分 43 秒 vs. 131 分 1 秒),两组辐射剂量无显著差异。

要点

  • 70 kVp ASIR-V 较 80 kVp MBIR 图像噪声更低、SNR 与 CNR 更高,且重建时间大幅缩短。
  • 两组辐射剂量无显著差异,临床信息亦无显著差异。
  • 除静脉轮廓勾画外,B 组主观图像质量指标均显著优于 A 组。

核心结论

  • 在辐射剂量相近的情况下,70 kVp ASIR-V 下肢 CT 静脉造影的图像质量和重建时间均优于 80 kVp MBIR。

诊断标准与关键数值

  • B 组血管强化、SNR 和 CNR 均显著高于 A 组(p ≤ 0.015)。
  • B 组图像噪声显著低于 A 组(p ≤ 0.021)。
  • 除静脉轮廓勾画外,B 组所有主观图像质量指标均显著优于 A 组(p ≤ 0.021)。
  • B 组平均重建时间 1 分 43 秒,显著短于 A 组 131 分 1 秒(p < 0.001)。
  • 两组临床信息和辐射剂量无显著差异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To compare the objective and subjective image quality indicators and radiation doses of computed tomography (CT) venography performed using model-based iterative reconstruction (MBIR) at 80 kVp and adaptive statistical iterative reconstruction (ASIR)-V at 70 kVp.

MATERIALS AND METHODS: Eighty-three patients who had undergone CT venography of the lower extremities with MBIR at 80 kVp (Group A; 21 men and 20 women; mean age, 55.5 years) or ASIR-V at 70 kVp (Group B; 18 men and 24 women; mean age, 57.3 years) were enrolled. Two radiologists retrospectively evaluated the objective (vascular enhancement, image noise, signal-to-noise ratio [SNR], contrast-to-noise ratio [CNR]) and subjective (quantum mottle, delineation of contour, venous enhancement) image quality indicators at the inferior vena cava and femoral and popliteal veins. Clinical information, radiation dose, reconstruction time, and objective and subjective image quality indicators were compared between groups A and B.

RESULTS: Vascular enhancement, SNR, and CNR were significantly greater in Group B than in Group A (p ≤ 0.015). Image noise was significantly lower in Group B (p ≤ 0.021), and all subjective image quality indicators, except for delineation of vein contours, were significantly better in Group B (p ≤ 0.021). Mean reconstruction time was significantly shorter in Group B than in Group A (1 min 43 s vs. 131 min 1 s; p < 0.001). Clinical information and radiation dose were not significantly different between the two groups.

CONCLUSION: CT venography using ASIR-V at 70 kVp was better than MBIR at 80 kVp in terms of image quality and reconstruction time at similar radiation doses.