文献速递神经外周血管CT

颈动脉CTA采用最低管电压(70 kV)与自动管电压适配方案的比较

Carotid CTA at the Lowest Tube Voltage (70 kV) in Comparison with Automated Tube Voltage Adaption.
PubMed 文献 · AJNR Am J Neuroradiol · 2019-07-11

前瞻性随机研究纳入90例连续患者,比较70 kV与自动管电压适配颈动脉CTA。70 kV组容积CT剂量指数降低22%、剂量长度乘积降低20%(各P < .001),对比度、信噪比、对比噪声比及主观图像质量与病理显示无显著差异,结论为70 kV在极低剂量水平下可行且整体图像质量与自动管电压选择方案相当。

要点

  • 70 kV颈动脉CTA方案容积CT剂量指数较自动管电压适配降低22%,剂量长度乘积降低20%。
  • 两组动脉血管对比度、信噪比、对比噪声比及参照辐射剂量的对比噪声比均无显著差异。
  • 两名观察者主观整体图像质量、伪影分析及诊断评估在两组间无显著差异。

核心结论

  • 颈动脉CTA采用可用的最低管电压(70 kV)在极低剂量水平下可行,整体图像质量与自动管电压选择方案相当。

研究设计与样本

  • 前瞻性随机研究纳入90例连续患者,随机分为研究组(n = 45,70 kV)和对照组(n = 45,自动管电压适配)。
  • 记录检查方案中的容积CT剂量指数和剂量长度乘积,图像质量评估包括动脉血管对比度、信噪比、对比噪声比及参照辐射剂量的对比噪声比。
  • 由2名观察者采用4分Likert量表进行主观整体图像质量分析、图像伪影分析和诊断评估。

关键影像表现/征象

  • 研究组辐射暴露显著更低:容积CT剂量指数降低22%,剂量长度乘积降低20%(各P < .001)。
  • 两组对比度(P = .15)、信噪比(P = .4)和对比噪声比(P = .5)无显著差异。
  • 参照辐射剂量的对比噪声比在研究方案下未显著增加(P = .2)。
  • 主观图像质量和病理发现显示在两组间无显著差异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: CTA is the imaging modality of choice in many institutions for the evaluation of the supraaortic vessels, but radiation exposure remains a matter of concern. Our aim was to evaluate a 70-kV protocol for CT angiography of the carotid arteries with respect to image quality and radiation exposure compared with automated tube voltage adaption.

MATERIALS AND METHODS: A total of 90 consecutive patients were included in this prospective study and randomized to the study group (n = 45, 70 kV) or control group (n = 45, automated tube voltage adaptation). Volume CT dose indices and dose-length products were recorded in the examination protocol. Image quality was assessed as arterial vessel contrast, signal-to-noise ratio, contrast-to-noise ratio, and contrast-to-noise ratio in reference to the radiation dose. Subjective overall image-quality analysis, image-artifact analysis, and diagnostic evaluation were performed by 2 observers by using a 4-point Likert scale.

RESULTS: Radiation exposure was significantly lower in the study group (volume CT dose index reduced by 22%, dose-length product reduction by 20%; each P < .001). Contrast (P = .15), SNR (P = .4), and contrast-to-noise ratio (P = .5) did not show significant differences between the groups. The contrast-to-noise ratio in reference to the radiation dose was not significantly increased using the study protocol (P = .2). Subjective image quality and visualization of pathologic findings did not differ significantly between the groups.

CONCLUSIONS: Carotid CTA using the lowest available voltage (70 kV) is feasible at very-low-dose levels, while overall image quality is comparable with protocols using automated tube voltage selection.

来源:AJNR Am J Neuroradiol,查看原文。版权归原作者所有。

引用

  1. Eller A, Wiesmüller M, Wüst W, et al. Carotid CTA at the Lowest Tube Voltage (70 kV) in Comparison with Automated Tube Voltage Adaption. AJNR Am J Neuroradiol. 2019. PMID: 31296525. doi:10.3174/ajnr.A6108

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。