猪脊柱解剖标本、CT图像与3D打印模型之间的误差测量
研究对10例猪腰椎进行CT扫描、解剖后再次扫描并3D打印,比较四种测量水平(首次CT、二次CT、解剖骨块、3D打印骨块),结果显示四种测量间无统计学差异,组内相关系数(3,1)=0.9553,3D打印模型与CT图像平均误差0.60 mm,与解剖骨块平均误差0.73 mm,提示3D打印模型能准确反映在体骨结构。
要点
- 四种测量水平(首次CT、二次CT、解剖骨块、3D打印骨块)间无统计学差异(单因素重复测量方差分析,F < F_crit,p值 > α = 0.05)。
- 组内相关系数(3,1)=0.9553,提示四种测量水平间及不同猪样本数据间可靠性良好。
- 3D打印模型与CT图像平均误差0.60 mm,与解剖骨块平均误差0.73 mm。
核心结论
- 使用本文所述流程制作的3D打印模型与CT图像平均误差为0.60 mm,与解剖骨块平均误差为0.73 mm。
- 3D打印模型能准确反映在体骨结构,为手术规划提供准确的3D印象。
研究设计与样本
- 研究使用10例猪腰椎脊柱椎体的CT扫描,随后进行解剖并再次扫描。
- 对CT图像进行处理以获得网格并进行3D打印,比较首次CT、二次CT、解剖骨块和3D打印骨块四种测量水平。
- 统计方法采用单因素重复测量方差分析和组内相关系数(3,1)。
诊断标准与关键数值
- 四种测量水平间无统计学差异(单因素重复测量方差分析,F < F_crit,p值 > α = 0.05)。
- 组内相关系数(3,1)=0.9553。
- 3D打印模型与CT图像平均误差0.60 mm,与解剖骨块平均误差0.73 mm。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: 3D printers are increasingly used in medical applications such as surgical planning, creation of implants and prostheses, and medical education. For the creation of reliable 3D printed models of the vertebral column, processing must be performed on CT images. This processing must be assessed and validated so that any error of the printed model can be recognized and minimized.
MATERIAL AND METHODS: In order to perform this validation, 10 CT scans of porcine lumbar spinal vertebra were used, which were then dissected and scanned again. CT image processing was performed to obtain a mesh and perform 3D printing.
RESULTS: There was no statistical difference among the four different levels of vertebrae measurements (first CT images, second CT images, anatomical piece of porcine bone and 3D printing of porcine bone; One Way repeated measure ANOVA, F < F_crit, p value > α = 0.05). The Intraclass Correlation also revealed a mean intraclass correlation coefficient (3,1) = 0.9553, which describes the reliability of all four levels in addition to the reliability of the data between porcine samples subjected to different levels of measurement. This shows that the average error is less than 1 mm.
CONCLUSIONS: The measurements of models created with 3D printers using the pipeline described in this paper have an average error of 0.60 mm with CT images and 0.73 mm with anatomical piece. Thus, 3D printed models accurately reflect in vivo bones and provide accurate 3D impressions to assist in surgical planning.