超声计算机辅助诊断系统对甲状腺结节的真实世界性能评估
本研究评估商用计算机辅助诊断系统(S-Detect 1 和 S-Detect 2)检测甲状腺癌的诊断性能。在 106 例患者的 218 个甲状腺结节中,S-Detect 1 的敏感度、特异度、阳性预测值、阴性预测值和准确率分别为 80.2%、82.6%、75.0%、86.3% 和 81.7%;S-Detect 2 分别为 81.4%、68.2%、62.5%、84.9% 和 73.4%。CAD 系统与放射科医师对钙化描述的观察者间一致性为一般(kappa=0.336),而 S-Detect 2 的最终诊断及各超声描述符显示中等至高度一致性。结论认为当前 CAD 系统诊断甲状腺癌的特异度有限,S-Detect 2 的主要局限之一是识别钙化不准确,需由放射科医师进行鉴别。
要点
- S-Detect 1 与 S-Detect 2 对甲状腺癌诊断敏感度相近(80.2% vs 81.4%),但 S-Detect 2 特异度较低(68.2% vs 82.6%)。
- CAD 系统与放射科医师对钙化描述的观察者间一致性仅为一般(kappa=0.336)。
- S-Detect 2 识别钙化不准确,需由放射科医师进行鉴别。
核心结论
- 当前 CAD 系统诊断甲状腺癌的特异度有限;S-Detect 2 的主要局限之一是识别钙化不准确,需由放射科医师进行鉴别。
研究设计与样本
- 研究论文;共纳入 106 例患者的 218 个甲状腺结节,评估商用 CAD 系统(S-Detect 1 和 S-Detect 2)检测甲状腺癌的诊断性能,以放射科医师诊断为参照。
关键影像表现/征象
- S-Detect 1 与放射科医师对钙化描述的观察者间一致性为一般(kappa=0.336)。
- S-Detect 2 的最终诊断及各超声描述符与放射科医师显示中等至高度一致性。
诊断标准与关键数值
- S-Detect 1 诊断甲状腺癌的敏感度 80.2%、特异度 82.6%、阳性预测值 75.0%、阴性预测值 86.3%、准确率 81.7%。
- S-Detect 2 诊断甲状腺癌的敏感度 81.4%、特异度 68.2%、阳性预测值 62.5%、阴性预测值 84.9%、准确率 73.4%。
临床应用与阅片要点
- CAD 系统诊断甲状腺癌的特异度有限,阅片时需结合放射科医师判断。
- S-Detect 2 识别钙化不准确,钙化的鉴别需由放射科医师进行。
局限性
- 当前 CAD 系统诊断甲状腺癌的特异度有限。
- S-Detect 2 的主要局限之一是识别钙化不准确,需由放射科医师进行鉴别。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
This study evaluated the diagnostic performance of a commercially available computer-aided diagnosis (CAD) system (S-Detect 1 and S-Detect 2 for thyroid) for detecting thyroid cancers. Among 218 thyroid nodules in 106 patients, the sensitivity, specificity, positive predictive value, negative predictive value and accuracy of the CAD systems were 80.2%, 82.6%, 75.0%, 86.3% and 81.7%, respectively, for the S-Detect 1 and 81.4%, 68.2%, 62.5%, 84.9% and 73.4%, respectively, for the S-Detect 2. The inter-observer agreement between the CAD system and radiologist for the description of calcifications was fair (kappa = 0.336), while the final diagnosis and each ultrasonographic descriptor showed moderate to substantial agreement for the S-Detect 2. To conclude, the current CAD systems had limited specificity in the diagnosis of thyroid cancer. One of the main limitations of the S-Detect 2 was its inaccuracy in recognizing calcifications, which meant that differentiation had to be undertaken by the radiologist.