文献速递乳腺技术与安全乳腺X线

使用新型美国放射学会数字乳腺X线摄影模体对现代数字乳腺X线摄影系统质量调查的影响:来自台湾地区全国性调查的证据

Impact of using the new American College of Radiology digital mammography phantom on quality survey in modern digital mammography systems: Evidence from nationwide surveys in Taiwan.
PubMed 文献 · Eur J Radiol · 2019-05-20

该研究在2017至2018年对239台数字乳腺X线摄影(DM)设备进行现场调查,比较美国放射学会(ACR)屏片乳腺X线摄影(SFM)模体与DM模体在评估模体图像质量和平均腺体剂量(AGD)方面的差异。结果显示,DM模体在纤维和斑点细节的加权模体评分及总加权评分上显著高于SFM模体,但总体通过率较低(DM模体84.1% vs SFM模体91.2%),主要归因于肿块通过率较低(84.5%)。结论认为ACR DM模体对DM系统中斑点和纤维的辨别更优,可作为实施DM质量控制项目及大规模调查的参考。

要点

  • DM模体总加权评分显著高于SFM模体(8.61±1.04 vs 8.23±0.77,p<0.0001),对斑点和纤维的辨别更精确。
  • DM模体总体通过率低于SFM模体(84.1% vs 91.2%),主要因肿块通过率较低(84.5%)。
  • 切换模体后AGD无显著差异(p=0.06)。

核心结论

  • ACR DM模体对DM系统中斑点和纤维的评估具有更好的辨别力,可检测微小差异。
  • 该研究可为实施DM质量控制项目及数字时代使用新DM模体开展大规模调查提供参考。

研究设计与样本

  • 研究类型为2017年和2018年开展的全国性现场调查,共调查239台DM设备。
  • 比较ACR屏片乳腺X线摄影(SFM)模体与DM模体在评估模体图像质量和平均腺体剂量(AGD)方面的差异。
  • 模体图像质量采用加权模体评分评估,评分考虑每个细节的大小。

诊断标准与关键数值

  • 从SFM模体切换到DM模体时,AGD无显著差异(p=0.06)。
  • DM模体在纤维和斑点方面的平均加权模体评分显著高于SFM模体,总加权模体评分亦显著更高(DM模体 vs SFM模体:8.61±1.04 vs 8.23±0.77,p<0.0001)。
  • DM模体总体通过率为84.1%,低于SFM模体的91.2%。
  • DM模体肿块通过率为84.5%,低于SFM模体,可解释总体通过率较低。

临床应用与阅片要点

  • 使用DM模体评估DM系统时,对斑点和纤维的辨别更精确,可检测微小差异。
  • DM模体总体通过率较低,主要因肿块通过率较低(84.5%),阅片和报告时需注意此差异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The purpose of this study was to investigate the impact of the new American College of Radiology (ACR) digital mammography (DM) phantom in evaluating phantom image quality (IQ) and average glandular dose (AGD) in a nationwide survey on DM systems.

METHODS: On-site surveys of 239 DM units were conducted in 2017 and 2018, and comparisons were made between ACR screen-film mammography (SFM) phantom and DM phantom for accessing phantom IQ and AGD. The phantom IQ was assessed using the weighted phantom score, considering the size of each detail.

RESULTS: When switching from SFM phantom to DM phantom, no significant difference was found in AGD (p = 0.06). The mean weighted phantom score was significantly higher for DM phantom than for SFM phantom in terms of fibers and specks, and so was the total weighted phantom score (DM phantom vs. SFM phantom: 8.61 ± 1.04 vs. 8.23 ± 0.77, p < 0.0001). The phantom IQ is thus more precise and can detect small differences when using DM phantom and investigating DM systems, especially for specks and fibers. However, the overall passing rate was lower for DM phantom (84.1%) than for SFM phantom (91.2%). This can be explained by the lower passing rate for mass (84.5%) with the DM phantom.

CONCLUSION: The ACR DM phantom provides better discernment to assess specks and fibers in DM systems. This study may serve as a reference for implementing a DM quality control program and when conducting large-scale surveys with the new DM phantom in the digital era.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Hwang YS, Tsai HY, Lin YY, et al. Impact of using the new American College of Radiology digital mammography phantom on quality survey in modern digital mammography systems: Evidence from nationwide surveys in Taiwan. Eur J Radiol. 2019. PMID: 31307658. doi:10.1016/j.ejrad.2019.05.014

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。