高帧率超声造影成像用于慢速淋巴流:超声压力和流速对微泡破坏及图像持续性的影响
本研究在淋巴管体模和兔腘窝淋巴结上评估高帧率超声造影检测慢速淋巴流,发现高帧率成像结合奇异值分解滤波可显著减少组织伪影,慢流速下图像对比度和信号持续性更易受超声振幅或机械指数变化影响,探头运动显著降低血管图像对比度,而运动校正可改善,兔淋巴结内流速小于3 mm/s的血管可清晰显示。
要点
- 高帧率超声造影结合奇异值分解滤波可在高临床频率下显著减少体模中的组织伪影。
- 慢流速下图像对比度和信号持续性更易受超声振幅或机械指数变化影响,可选择机械指数值在图像对比度与微泡破坏之间取得折中。
- 探头运动显著降低血管图像对比度,奇异值分解滤波前进行运动校正可改善。
核心结论
- 高帧率超声造影成像结合奇异值分解滤波可用于检测慢速淋巴流,超声压力和流速影响微泡破坏与图像持续性,探头运动需运动校正。
研究设计与样本
- 在淋巴管体模上评估高帧率超声造影检测慢速淋巴流,并进一步在兔腘窝淋巴结血管上进行实验。
关键影像表现/征象
- 高帧率成像和奇异值分解滤波可在高临床频率下显著减少体模中的组织伪影。
- 慢流速下图像对比度和信号持续性更易受超声振幅或机械指数变化影响。
- 探头运动显著降低血管图像对比度,奇异值分解滤波前应用运动校正可改善。
- 兔淋巴结内血流速度小于3 mm/s的血管可清晰显示。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Contrast-enhanced ultrasound (CEUS) utilising microbubbles shows great potential for visualising lymphatic vessels and identifying sentinel lymph nodes (SLNs) which are valuable for axillary staging in breast cancer patients. However, current CEUS imaging techniques have limitations that affect the accurate visualisation and tracking of lymphatic vessels and SLN. (i) Tissue artefacts and bubble disruption can reduce the image contrast. (ii) Limited spatial and temporal resolution diminishes the amount of information that can be captured by CEUS. (iii) The slow lymph flow makes Doppler-based approaches less effective. This work evaluates on a lymphatic vessel phantom the use of high frame rate (HFR) CEUS for the detection of lymphatic vessels where flow is slow. Specifically, the work particularly investigates the impact of key factors in lymphatic imaging, including ultrasound pressure and flow velocity as well as probe motion during vessel tracking, on bubble disruption and image contrast. Experiments were also conducted to apply HFR CEUS imaging on vasculature in a rabbit popliteal lymph node (LN). Our results show that (i) HFR imaging and singular value decomposition (SVD) filtering can significantly reduce tissue artefacts in the phantom at high clinical frequencies; (ii) the slow flow rate within the phantom makes image contrast and signal persistence more susceptible to changes in ultrasound amplitude or mechanical index (MI), and an MI value can be chosen to reach a compromise between images contrast and bubble disruption under slow flow condition; (iii) probe motion significantly decreases image contrast of the vessel, which can be improved by applying motion correction before SVD filtering; (iv) the optical observation of the impact of ultrasound pressure on HFR CEUS further confirms the importance of optimising ultrasound amplitude and (v) vessels inside rabbit LN with blood flow less than 3 mm/s are clearly visualised.
引用
- Zhu J, Lin S, Leow CH, et al. High Frame Rate Contrast-Enhanced Ultrasound Imaging for Slow Lymphatic Flow: Influence of Ultrasound Pressure and Flow Rate on Bubble Disruption and Image Persistence. Ultrasound Med Biol. 2019. PMID: 31279503. doi:10.1016/j.ultrasmedbio.2019.05.016
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。