低b值(50-100)扩散加权成像在儿童中检出的高信号肝脏病灶数显著多于T2加权成像
该回顾性研究纳入50例儿童肝脏MRI,以全序列共识阅片结合随访、其他影像和病理为参考标准,比较低b值DWI与T2加权序列对高信号局灶性肝脏病灶的检出敏感度。低b值DWI平均检出134个(79%),T2加权检出95个(56%),参考标准共170个病灶;低b值DWI较T2加权检出更多病灶。
要点
- 儿童肝脏MRI中,低b值DWI对高信号局灶性病灶的检出敏感度高于T2加权序列。
- 低b值DWI的观察者间一致性ICC为0.96,T2加权为0.85,低b值DWI略优。
- 两种序列在识别最小病灶大小上无显著差异。
核心结论
- 低b值DWI图像在检出儿童高信号局灶性肝脏病灶方面比T2加权序列更敏感。
研究设计与样本
- 回顾性研究,纳入50例儿童(22例男性、28例女性;年龄2个月至17岁,平均10.9岁)的肝脏MRI,评估局灶性肝脏病灶。
- 两名放射科医生在不同时间独立阅片低b值DWI和T2加权序列,记录病灶、最小病灶大小和位置。
- 以全部MRI序列的共识阅片并结合随访、其他影像学检查和可获得病例的病理结果作为参考标准,确定最终病灶数。
关键影像表现/征象
- 共识阅片确定170个高信号病灶作为参考标准。
- 两名阅片者低b值DWI平均检出134个(79%)病灶,T2加权平均检出95个(56%)病灶。
- 低b值DWI检出病灶的观察者间一致性ICC=0.96(0.93-0.98),T2加权ICC=0.85(0.75-0.91),低b值DWI一致性略好。
诊断标准与关键数值
- 与参考标准(平均病灶数3.45)相比,T2加权检出的病灶数显著更少(平均病灶数1.91;P=0.0001),低b值DWI检出病灶数与参考标准无显著差异(平均病灶数2.68;P=0.1527)。
- 两名阅片者在低b值DWI上检出的病灶数均显著多于T2加权(阅片者1:P=0.0036;阅片者2:P=0.0001)。
- 低b值DWI与T2加权在识别最小病灶大小上无显著差异(阅片者1:P=0.19;阅片者2:P=0.47)。
临床应用与阅片要点
- 儿童肝脏MRI检出高信号局灶性病灶时,低b值DWI较T2加权序列更敏感,阅片时可结合两者以提高病灶检出。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Low b-value (50-100 s/mm2) diffusion-weighted images (low b-value DWI) have been shown to be superior to T2-weighted fast spin echo sequence (T2-W) in detecting liver lesions in adults. There are no such studies assessing this difference in children.
OBJECTIVE: The purpose of the study was to compare the sensitivity of low b-value DWI images and T2-W images in detecting focal liver lesions in children.
MATERIALS AND METHODS: A retrospective review of liver magnetic resonance imaging (MRI) to assess focal liver lesions in 50 children (22 males, 28 females; age: 2 months to 17 years [mean: 10.9 years]) was performed. Two radiologists reviewed both low b-value DWI and T2-W sequences independently on different occasions to note lesions, the smallest lesion size and the location. A consensus reading of the entire MRI examination and a correlation with follow-up, other imaging modalities and pathology in available cases were used to determine the final number of lesions as a reference standard. The average number of lesions per patient detected by both readers on each sequence was compared with each other and with the reference standard using the signed-rank test. The smallest lesions detected by each sequence were compared using the paired t-test.
RESULTS: One hundred seventy hyperintense lesions were identified on consensus review to serve as the reference standard. The average number of lesions identified by both readers on low b-value DWI was 134 (79%) and on T2-W was 95 (56%). There was excellent interobserver agreement for detection of lesions on low b-value DWI (intraclass correlation coefficient [ICC]=0.96 [0.93-0.98]) and T2-W (ICC=0.85 [0.75-0.91]), with slightly better agreement on low b-value DWI. Both readers identified significantly more number of lesions on low b-value DWI compared to T2-W (Reader 1: P=0.0036, Reader 2: P=0.0001). Compared to the reference standard (mean number of lesions: 3.45), T2-W detected significantly fewer lesions (mean number of lesions: 1.91; P=0.0001) while there was no significant difference in lesion detection on low b-value DWI (mean number of lesions: 2.68; P=0.1527). Low b-value DWI and T2-W were not significantly different in identifying the smallest lesion size (Reader 1: P=0.19, Reader 2: P=0.47).
CONCLUSION: Low b-value DWI images are more sensitive than T2-W sequences in detecting hyperintense focal liver lesions in children.
引用
- Grasparil ADI, Gupta H, Sheybani E, et al. Low b-value (50-100) diffusion-weighted images detect significantly more hyperintense liver lesions in children than T2-weighted images. Pediatr Radiol. 2019. PMID: 31292681. doi:10.1007/s00247-019-04455-9
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。