脉冲与连续高强度聚焦超声消融的体外与体内模型比较研究
该研究在相同剂量(1200 J)下比较脉冲高强度聚焦超声(PHIFU,占空比60%和20%)与连续高强度聚焦超声(CHIFU)消融牛肝(体外)和兔肝(体内)的效果。结果显示PHIFU产生更多液化性坏死,60%占空比PHIFU能量效率因子显著低于CHIFU(P < .05),CHIFU灰度值和平均灰度密度显著高于PHIFU(P < .05)。作者认为单次PHIFU可产生液化性坏死,较高占空比可提高消融效率。
要点
- 相同剂量下,PHIFU各组均出现液化性坏死,而CHIFU未见。
- 60%占空比PHIFU的能量效率因子在体外和体内均显著低于CHIFU(P < .05)。
- CHIFU的灰度值和平均灰度密度显著高于PHIFU(60%或20%;P < .05)。
核心结论
- 相同剂量下,与CHIFU相比,单次PHIFU可产生液化性坏死,且较高占空比可提高消融效率;这种增效可能涉及空化效应增强组织破坏以及通过空化效应或更有效热扩散减少内源性微气泡的阻碍效应。
研究设计与样本
- 体外研究使用牛肝组织,体内研究使用兔肝组织,比较连续HIFU与脉冲HIFU(1200 J,占空比60%和20%)的消融能力;消融后观察灰度变化和病理特征,计算组织坏死体积、能量效率因子和平均灰度密度。
关键影像表现/征象
- 脉冲模式产生更明显的液化性坏死。
- 部分20%占空比PHIFU样本灰度变化不明显,表现为椭圆形腔隙。
- CHIFU引起的灰度值和平均灰度密度显著高于PHIFU(60%或20%;P < .05)。
诊断标准与关键数值
- PHIFU在60%占空比下的能量效率因子在体外和体内均显著低于CHIFU(P < .05)。
- CHIFU的灰度值和平均灰度密度显著高于PHIFU(60%或20%;P < .05)。
- 组织病理学分析显示所有PHIFU组均存在液化性坏死。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To compare the efficacy of pulsed high-intensity focused ultrasound (PHIFU) versus continuous high-intensity focused ultrasound (CHIFU) ablation at identical doses.
METHODS: Continuous and pulsed HIFU (1200 J) at duty cycles (DCs) of 60% and 20% were examined for their capacity to ablate bovine liver tissue in vitro and rabbit liver tissue in vivo. After ablation, grayscale changes and pathologic characteristics were observed or measured, and the tissue necrosis volume, energy efficiency factor, and average grayscale density were calculated.
RESULTS: The pulsed mode generated greater liquefaction necrosis. An inconspicuous grayscale change was observed for PHIFU at a DC of 20% in some samples, which appeared as an elliptical cavity. The energy efficiency factor of PHIFU at a DC of 60% was significantly lower than that of CHIFU, as observed both in vitro and in vivo (P < .05). The grayscale value and average grayscale density in response to CHIFU were significantly greater than those in response to PHIFU (60% or 20%; P < .05). Histopathologic analysis revealed liquefaction necrosis in all PHIFU groups.
CONCLUSIONS: At identical doses, compared with CHIFU, a single session of PHIFU can generate liquefaction necrosis and at a higher DC can improve ablation efficiency. This increased efficacy of PHIFU may involve enhancement of tissue destruction by cavitation effects and a reduction in the obstruction effect of endogenous microbubbles through cavitation effects or a more effective diffusion of heat.
引用
- Wang Y, Wang Q, Luo Y, et al. Comparative Study of Pulsed Versus Continuous High-Intensity Focused Ultrasound Ablation Using In Vitro and In Vivo Models. J Ultrasound Med. 2019. PMID: 31339599. doi:10.1002/jum.15098
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。