CT迭代重建算法的基于任务图像质量评估
该研究使用ACR体模在五个剂量水平下,比较四家CT厂商的滤波反投影与两代迭代重建算法的图像质量。通过噪声功率谱和任务传递函数计算可探测性指数,模拟肝脏大肿块和小的钙化灶检测任务,评估不同算法的潜在剂量降低能力。结果表明基于任务的方法可稳健用于优化过程。
要点
- 可探测性指数d'适用于CT迭代重建算法的剂量优化评估
- 混合/统计迭代重建中,Siemens的d'最高,Canon次之,Philips和GE最低
- 基于模型的迭代重建对GE、Philips、Siemens可分别降低大肿块检测剂量35%、62%、13%
核心结论
- 多厂商比较迭代重建算法可评估各厂商算法演进,d'高度适用于优化过程。
研究设计与样本
- 使用四套CT系统(配备混合/统计迭代重建和全/部分/高级基于模型迭代重建算法),在ACR体模上五个剂量水平采集,原始数据用标准软组织核重建,计算噪声功率谱和任务传递函数。
诊断标准与关键数值
- 大肿块特征:120 HU,直径25 mm;小钙化特征:500 HU,直径1.5 mm
- 混合/统计迭代重建下,大肿块d'最高为Siemens,其次Canon,最低Philips和GE;小钙化d'排序相同
- 基于模型迭代重建相比混合/统计迭代重建的潜在剂量降低:大肿块GE -35%、Philips -62%、Siemens -13%;小钙化GE -45%、Philips -78%、Siemens -14%
- Canon系统基于模型迭代重建相比混合/统计迭代重建,仅小钙化潜在剂量降低-32%,大肿块剂量增加100%
临床应用与阅片要点
- 基于任务的图像质量评估可提供临床有意义的图像质量指标,辅助CT迭代重建算法优化。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To assess the dose performance in terms of image quality of filtered back projection (FBP) and two generations of iterative reconstruction (IR) algorithms developed by the most common CT vendors.
MATERIALS AND METHODS: We used four CT systems equipped with a hybrid/statistical IR (H/SIR) and a full/partial/advanced model-based IR (MBIR) algorithms. Acquisitions were performed on an ACR phantom at five dose levels. Raw data were reconstructed using a standard soft tissue kernel for FBP and one iterative level of the two IR algorithm generations. The noise power spectrum (NPS) and the task-based transfer function (TTF) were computed. A detectability index (d') was computed to model the detection task of a large mass in the liver (large feature; 120 HU and 25-mm diameter) and a small calcification (small feature; 500 HU and 1.5-mm diameter).
RESULTS: With H/SIR, the highest values of d' for both features were found for Siemens, then for Canon and the lowest values for Philips and GE. For the large feature, potential dose reductions with MBIR compared with H/SIR were - 35% for GE, - 62% for Philips, and - 13% for Siemens; for the small feature, corresponding reductions were - 45%, - 78%, and - 14%, respectively. With the Canon system, a potential dose reduction of - 32% was observed only for the small feature with MBIR compared with the H/SIR algorithm. For the large feature, the dose increased by 100%.
CONCLUSION: This multivendor comparison of several versions of IR algorithms allowed to compare the different evolution within each vendor. The use of d' is highly adapted and robust for an optimization process.
KEY POINTS: • The performance of four CT systems was evaluated by using imQuest software to assess noise characteristic, spatial resolution, and lesion detection. • Two task functions were defined to model the detection task of a large mass in the liver and a small calcification. • The advantage of task-based image quality assessment for radiologists is that it does not include only complicated metrics, but also clinically meaningful image quality.