模型迭代重建与滤过反投影及混合迭代重建在不同剂量下对部分实性结节和实性结节CT体积测量准确性的比较:体模研究
该体模研究在五种辐射剂量下,比较模型迭代重建(MIR)、滤过反投影(FBP)和混合迭代重建(HIR)对部分实性结节(PSN)和实性结节(SN)体积测量的准确性。结果显示,MIR在全部结节和剂量条件下绝对百分比测量误差(APE)均显著低于FBP和HIR(均p<0.01),且对最小PSN内部实性部分(3 mm)和最小SN(3 mm)的APE最低。MIR图像噪声在所有剂量下也显著低于FBP和HIR(均p<0.001)。
要点
- MIR在所有剂量下对PSN和SN的体积测量误差均显著低于FBP和HIR(均p<0.01)。
- MIR对最小PSN内部实性部分(3 mm)和最小SN(3 mm)的APE最低。
- MIR图像噪声在所有剂量下均显著低于FBP和HIR(均p<0.001)。
核心结论
- MIR在低剂量与标准剂量条件下对PSN和SN的体积测量准确性均优于FBP和HIR,对最小PSN和SN的体积测量有效。
研究设计与样本
- 离体胸部体模研究,将8种不同直径的PSN和SN置入体模,在5种辐射剂量水平(120 kVp/100 mAs、120 kVp/50 mAs、120 kVp/20 mAs、120 kVp/10 mAs、80 kVp/10 mAs)下进行CT扫描。
- 每次扫描分别用FBP、HIR或MIR的3种图像定义(IMR-R1、IMR-ST1、IMR-SP1;Philips Healthcare)重建图像,半自动测量SN和PSN(含实性/磨玻璃部分)体积并计算绝对百分比测量误差(APE)。
诊断标准与关键数值
- APE(绝对百分比测量误差)用于评价体积测量准确性,图像噪声用于评价图像质量。
- 所有结节和剂量条件下MIR的APE均显著低于FBP和HIR(均p<0.01)。
- 最小PSN内部实性部分(3 mm)和SN(3 mm)的APE在MIR(IMR-R1和IMR-ST1)重建时最低:120 kVp/100 mAs为1.06 ± 1.36和8.75 ± 3.96(p<0.001);120 kVp/50 mAs为1.95 ± 1.56和5.61 ± 0.85(p=0.002);120 kVp/20 mAs为2.88 ± 3.68和5.75 ± 1.95(p=0.001);120 kVp/10 mAs为5.57 ± 6.26和6.32 ± 2.91(p=0.091);80 kVp/10 mAs为5.84 ± 1.96和6.90 ± 3.31(p=0.632)。
- MIR图像噪声在所有剂量下均显著低于FBP和HIR(均p<0.001):120 kVp/100 mAs为3.22 ± 0.66;120 kVp/50 mAs为4.19 ± 1.37;120 kVp/20 mAs为5.49 ± 1.16;120 kVp/10 mAs为6.88 ± 1.91;80 kVp/10 mAs为12.49 ± 6.14。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To investigate the accuracy of model-based iterative reconstruction (MIR) for volume measurement of part-solid nodules (PSNs) and solid nodules (SNs) in comparison with filtered back projection (FBP) or hybrid iterative reconstruction (HIR) at various radiation dose settings.
MATERIALS AND METHODS: CT scanning was performed for eight different diameters of PSNs and SNs placed in the phantom at five radiation dose levels (120 kVp/100 mAs, 120 kVp/50 mAs, 120 kVp/20 mAs, 120 kVp/10 mAs, and 80 kVp/10 mAs). Each CT scan was reconstructed using FBP, HIR, or MIR with three different image definitions (body routine level 1 [IMR-R1], body soft tissue level 1 [IMR-ST1], and sharp plus level 1 [IMR-SP1]; Philips Healthcare). The SN and PSN volumes including each solid/ground-glass opacity portion were measured semi-automatically, after which absolute percentage measurement errors (APEs) of the measured volumes were calculated. Image noise was calculated to assess the image quality.
RESULTS: Across all nodules and dose settings, the APEs were significantly lower in MIR than in FBP and HIR (all p < 0.01). The APEs of the smallest inner solid portion of the PSNs (3 mm) and SNs (3 mm) were the lowest when MIR (IMR-R1 and IMR-ST1) was used for reconstruction for all radiation dose settings. (IMR-R1 and IMR-ST1 at 120 kVp/100 mAs, 1.06 ± 1.36 and 8.75 ± 3.96, p < 0.001; at 120 kVp/50 mAs, 1.95 ± 1.56 and 5.61 ± 0.85, p = 0.002; at 120 kVp/20 mAs, 2.88 ± 3.68 and 5.75 ± 1.95, p = 0.001; at 120 kVp/10 mAs, 5.57 ± 6.26 and 6.32 ± 2.91, p = 0.091; at 80 kVp/10 mAs, 5.84 ± 1.96 and 6.90 ± 3.31, p = 0.632). Image noise was significantly lower in MIR than in FBP and HIR for all radiation dose settings (120 kVp/100 mAs, 3.22 ± 0.66; 120 kVp/50 mAs, 4.19 ± 1.37; 120 kVp/20 mAs, 5.49 ± 1.16; 120 kVp/10 mAs, 6.88 ± 1.91; 80 kVp/10 mAs, 12.49 ± 6.14; all p < 0.001).
CONCLUSION: MIR was the most accurate algorithm for volume measurements of both PSNs and SNs in comparison with FBP and HIR at low-dose as well as standard-dose settings. Specifically, MIR was effective in the volume measurement of the smallest PSNs and SNs.
引用
- Kim SK, Kim C, Lee KY, et al. Accuracy of Model-Based Iterative Reconstruction for CT Volumetry of Part-Solid Nodules and Solid Nodules in Comparison with Filtered Back Projection and Hybrid Iterative Reconstruction at Various Dose Settings: An Anthropomorphic Chest Phantom Study. Korean J Radiol. 2019. PMID: 31270983. doi:10.3348/kjr.2018.0893