双能CT头颈部成像中生理盐水冲管技术对虚拟单能图像质量及静脉周围伪影的影响
本研究将50例头颈癌患者分为无冲管组(A组)与30ml生理盐水冲管组(B组),行快速千伏切换双能CT并重建虚拟单能图像。B组在40-80keV各能级病灶对比噪声比均更高,最佳能级A组63keV、B组61keV;B组静脉周围伪影评分显著更高(40keV时2.44对2.74,最佳能级时3.20对3.46,均p<0.05),主观病灶显示相当。结论为冲管技术可改善病灶CNR并减少静脉周围伪影,尤其在40keV。
要点
- 生理盐水冲管组在40-80keV各能级病灶CNR均高于无冲管组。
- 冲管组最佳虚拟单能能级为61keV,无冲管组为63keV。
- 冲管组静脉周围伪影评分显著更高(40keV时2.44对2.74,最佳能级时3.20对3.46,均p<0.05)。
核心结论
- 生理盐水冲管技术可改善双能CT虚拟单能成像的病灶对比噪声比并减少静脉周围伪影,尤其在40keV时。
研究设计与样本
- 研究纳入50例头颈癌患者,分为无冲管组(A组)与30ml生理盐水冲管组(B组)。
- 所有图像采用快速千伏切换CT(Revolution HD,GE Healthcare,Milwaukee,WI,USA)采集,并重建虚拟单能图像。
- 在40-80keV能级计算病灶对比噪声比,由两名 reviewer 在40keV和客观分析确定的最佳能级进行主观图像质量、病灶显示及静脉周围伪影评分。
关键影像表现/征象
- 最佳虚拟单能能级A组为63keV,B组为61keV。
- 在40-80keV能级范围内,B组病灶对比噪声比高于A组。
- B组在最佳能级的主观整体图像质量最高(主观图像质量均值3.40)。
- 两组主观病灶显示相当。
- B组静脉周围伪影评分显著更高:40keV时2.44对2.74(p<0.05),最佳能级时3.20对3.46(p<0.05)。
诊断标准与关键数值
- 病灶对比噪声比在虚拟单能40-80keV能级计算,最佳能级由客观分析确定(A组63keV,B组61keV)。
- 主观整体图像质量、病灶显示及静脉周围伪影由两名 reviewer 在40keV和最佳能级评分,B组最佳能级主观整体图像质量均值3.40。
- 静脉周围伪影评分B组显著高于A组:40keV时2.44对2.74(p<0.05),最佳能级时3.20对3.46(p<0.05)。
临床应用与阅片要点
- 原文结论指出生理盐水冲管技术可改善双能CT虚拟单能成像的病灶对比噪声比并减少静脉周围伪影,尤其在40keV时。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To evaluate the effect of the saline flush (SF) technique on the depiction of lesions and the reduction of perivenous artefacts in the head and neck region using dual-energy computed tomography (CT) with virtual monochromatic imaging (VMI).
MATERIALS AND METHODS: Fifty patients with head and neck cancer were divided into two groups: group A, without a SF and group B, with a 30-ml SF. All images were acquired using fast kilovolt-switching CT (Revolution HD, GE Healthcare, Milwaukee, WI, USA). Contrast-to-noise ratios (CNRs) of the lesions were calculated at VMI energy levels ranging from 40 to 80 keV. Subjective analysis of overall image quality, delineation of lesions, and perivenous artefacts was conducted by two reviewers at both VMI energy level 40 keV and the optimal energy level (which showed optimal CNR by objective analysis).
RESULTS: Optimal energy level was 63 keV for group A and 61 keV for group B. At VMI energy levels ranging from 40 to 80 keV, the CNR was higher for group B. The highest subjective overall image quality was shown for group B at the optimal energy level (subjective image quality mean value, 3.40). Subjective delineation of lesions was comparable. The perivenous artefact score was significantly higher for group B (2.44 versus 2.74 [p<0.05] at 40 keV, 3.20 versus 3.46 [p<0.05] at the optimal energy level).
CONCLUSION: The SF technique results in an improvement of lesion CNR and a reduction of perivenous artefacts in VMI using duel-energy CT, especially at 40 keV.