儿童牙科与颌面部CBCT辐射剂量的日益关注:日常实践简易指南
该研究用EGSnrc蒙特卡洛代码模拟5种商用CBCT设备,结合头颈部体素模型评估5至14岁儿童各临床协议下器官剂量与癌症终生归因风险。女性LAR均高于男性,剂量随曝光年龄增长而降低,上中切牙协议风险最低、颅骨协议最高,唾液腺和口腔黏膜受照最多。
要点
- 女性癌症终生归因风险在所有协议中均高于男性。
- 曝光年龄越大,器官剂量与风险呈降低趋势。
- 扫描视野越小,器官剂量与辐射风险越低。
核心结论
- 该工作为5种牙科CBCT提供了详尽的剂量评估指南,可对每名儿科患者进行详细剂量评估。
研究设计与样本
- 在EGSnrc蒙特卡洛代码中模拟5种商用CBCT设备,使用自建头颈部体素模型(每个含22个分割器官),评估5至14岁男女各临床协议的器官剂量与癌症终生归因风险。
关键影像表现/征象
- 唾液腺和口腔黏膜在所有协议中均为受照最高的器官。
- 整个鼻腔位于主射野内的协议中,胸外组织(ET)为次高受照器官。
- 年幼患者甲状腺剂量相当高。
临床应用与阅片要点
- 作者指出,儿科牙科与颌面部CBCT使用增多带来的辐射剂量担忧,凸显了正当性判断的必要性,该判断应部分基于放射学中的辐射暴露。
- 作者指出,基于蒙特卡洛模拟和头颈部儿科体素模型的患者特异性剂量计算克服了传统热释光剂量计(TLD)剂量学的局限,并为CBCT曝光的正当性判断提供了适当指导。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To provide an indication-based and scanner-specific radiation dose and risk guide for paediatric patients undergoing dental and maxillofacial cone beam computed tomography (CBCT) examinations.
METHODS: Five commercially available scanners were simulated in EGSnrc Monte Carlo (MC) code. Dedicated, in-house built, head and neck voxel models, each consisting of 22 segmented organs, were used in the study. Organ doses and life attributable risk (LAR) for cancer incidence were assessed for males and females, aged 5 to 14 years old, for every clinically available protocol: central upper and lower incisors, upper and lower premolars, upper and lower jaws, cleft palate, temporal bone, sinus, dentomaxillofacial complex, and face and skull imaging. Dose results were normalised to the x-ray tube load (mAs) and logarithmic curves were fit to organ dose and risk versus age data.
RESULTS: Females demonstrated higher LAR values in all cases. A well-established dose decreasing pattern with increasing age-at-exposure was observed. Central upper incisor protocols were those with the lowest risk, contrary to skull protocols which provided the highest LAR values. Salivary glands and oral mucosa were the highest irradiated organs in all cases, followed by extrathoracic tissue (ET) in protocols where the entire nasal cavity was inside the primary field. The dose to thyroid was considerably high for younger patients.
CONCLUSIONS: This work provides an extensive dose assessment guide for 5 dental CBCTs, enabling detailed dose assessment for every paediatric patient.
KEY POINTS: • Radiation dose concerns due to the growing use of paediatric dental and maxillofacial CBCT underline the need for justification that should in part be based on radiation exposure in radiology. • Patient-specific dose calculations based on Monte Carlo simulations and head-neck paediatric voxel models overcome the limitations of conventional thermoluminescent dosimeter (TLD) dosimetry and provide proper guidance for justification of CBCT exposures. • Monte Carlo simulations with head-neck models reveal an organ dose and radiation risk decreasing pattern with increasing age at exposure, and with decreasing size of the scanning volume of interest (field of view).
引用
- Stratis A, Zhang G, Jacobs R, et al. The growing concern of radiation dose in paediatric dental and maxillofacial CBCT: an easy guide for daily practice. Eur Radiol. 2019. PMID: 31264018. doi:10.1007/s00330-019-06287-5
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。