患者偏离中心对儿童头部及胸腹部CT器官剂量和图像噪声的影响

How patient off-centering impacts organ dose and image noise in pediatric head and thoracoabdominal CT.
PubMed 文献 · Eur Radiol · 2019-07-05

该研究使用模拟5岁儿童的人体模型,在垂直床位-6至+6 cm范围内重复行头部及胸腹部CT,评估偏离中心对器官剂量和图像噪声的影响。结果显示器官剂量差异最高达34%,头部CT主要受蝴蝶结滤器影响,胸腹部CT受自动管电流调制和蝴蝶结滤器共同影响,散射辐射与体位无显著相关(p=0.21)。

要点

  • 儿童CT垂直体位偏离中心可使器官剂量差异最高达34%,头部和胸腹部CT的影响规律不同。
  • 头部CT中剂量与体位呈线性关系,主要受蝴蝶结滤器影响。
  • 胸腹部CT中剂量与体位呈二次方关系,受自动管电流调制和蝴蝶结滤器共同影响。

核心结论

  • 在儿童CT中,垂直体位对辐射剂量有实质性影响,差异最高达34%,取决于身体区域和每个器官的位置。

研究设计与样本

  • 使用模拟5岁儿童的人体模型,在垂直床位-6至+6 cm范围内重复行头部及胸腹部CT,采用自动管电流调制和默认蝴蝶结滤器,记录管电流时间乘积、器官剂量和图像噪声。

关键影像表现/征象

  • 与0位置相比,幕上脑剂量差异最高达22%,幕下脑最高达34%,眼最高达19%,肺最高达28%,胃最高达25%,肝最高达22%。
  • 头部CT中体位与剂量呈线性关系,主要受蝴蝶结滤器影响。
  • 胸腹部CT中体位与剂量呈二次方关系,受自动管电流调制和蝴蝶结滤器影响。
  • 剂量差异还取决于每个器官相对于等中心的相对位置。
  • 体位与图像噪声呈反向关系。

诊断标准与关键数值

  • 头部CT中甲状腺散射辐射与患者体位无显著相关(p=0.21)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To assess the impact of patient off-centering on organ dose and image noise for head and thoracoabdominal CT in a pediatric phantom.

METHODS: An anthropomorphic phantom simulating a 5-year-old child was used. Semiconductor dosimeters were placed in various cranial and thoracoabdominal organs. Head and thoracoabdominal CT were performed using automatic tube current modulation (ATCM) and default bowtie filters. The phantom was imaged repeatedly at vertical table positions ranging from - 6 to + 6 cm from the 0-position. Tube current time products (TCTP), organ doses, and image noise were recorded. Scatter radiation was measured in the thyroid for head CT. The effect of ATCM and bowtie filters was assessed.

RESULTS: Depending on patient position, organ doses differed up to 22% for the supratentorial brain, 34% for the infratentorial brain, 19% for the eyes, 28% for the lungs, 25% for the stomach, and 22% for the liver compared with those in the 0-position. The relation between position and dose was linear and mainly affected by the bowtie filter in head CT, while it was quadratic and affected by ATCM and bowtie filter in thoracoabdominal CT. It further depended on the relative position of each organ to the isocenter. An inverse relation was found between position and image noise. Scatter radiation was not significantly related to patient positioning (p = 0.21).

CONCLUSIONS: In pediatric CT, vertical patient positioning had a substantial impact on radiation dose with differences of up to 34%, depending on the body region and location of each individual organ.

KEY POINTS: • Patient off-centering has a substantial impact on organ radiation dose and image noise in pediatric CT. • Impact of patient off-centering on radiation dose and noise differs between head and thoracoabdominal CT. • Differences are caused by both ATCM and bowtie filter in thoracoabdominal CT, but mainly by bowtie filter in head CT.

来源:Eur Radiol,查看原文。版权归原作者所有。