指南与共识神经技术与安全MR

3.0-T磁共振成像伪影意外小于1.5 T:推荐对磁共振兼容可调脑室腹腔分流装置患者使用3.0-T扫描仪

Unexpectedly Smaller Artifacts of 3.0-T Magnetic Resonance Imaging than 1.5 T: Recommendation of 3.0-T Scanners for Patients with Magnet-Resistant Adjustable Ventriculoperitoneal Shunt Devices.
指南 / 共识 · World Neurosurg · 2019-06-28

该研究通过2名志愿者和4例患者比较3.0-T与1.5-T扫描仪对4种磁共振兼容可调分流装置的伪影大小,发现3.0-T在DWI和MRA上伪影更小,GE 3.0-T在所有序列上伪影均更小,4例临床病例结果一致。结论推荐对磁共振兼容可调分流装置患者,在诊断急性缺血性疾病或使用GE扫描仪时优先选择3.0-T扫描仪。

要点

  • 对磁共振兼容可调分流装置,3.0-T扫描仪在DWI和MRA上的伪影小于1.5-T扫描仪。
  • GE 3.0-T扫描仪在所有序列上伪影均小于1.5-T扫描仪,其他序列结果取决于制造商。
  • 4例临床病例的结果与体外实验结果一致。

核心结论

  • 对磁共振兼容可调分流装置患者,在诊断急性缺血性疾病或使用GE扫描仪时推荐使用3.0-T扫描仪而非1.5-T扫描仪。

研究设计与样本

  • 经机构伦理委员会批准,2名志愿者使用GE、Siemens和Philips的3.0-T与1.5-T扫描仪,将4种磁共振兼容可调分流装置(proGAV2.0、Codman Certas Plus、Polaris、Strata MR valve)固定在左侧颞部头皮,行常规MRI序列(T1WI、T2WI、FLAIR、DWI、MRA),并比较4例患者中3.0-T与1.5-T扫描仪的伪影。

关键影像表现/征象

  • 在DWI和MRA图像上,所有分流装置和所有扫描仪的3.0-T扫描仪伪影均小于1.5-T扫描仪。
  • 在其他序列上,伪影大小取决于MRI扫描仪制造商,但GE 3.0-T扫描仪在所有序列上伪影均更小。
  • 4例临床病例中,3.0-T扫描仪伪影小于1.5-T扫描仪的结果同样成立。

临床应用与阅片要点

  • 对磁共振兼容可调分流装置患者,在诊断急性缺血性疾病或使用GE扫描仪时,原文推荐优先选择3.0-T扫描仪而非1.5-T扫描仪。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Magnetic resonance imaging (MRI) artifacts of adjustable shunt devices are thought to be similar to metal clip artifacts, in that they are larger with higher field strength scanners. We have published several reports about the artifacts of new MRI-resistant adjustable shunt devices, and we found a case in which a 3.0-T scanner showed smaller artifacts than the 1.5-T scanner. We aimed to clarify whether this claim is true or not.

METHODS: Under permission of our institutional Ethical Committee, 2 volunteers underwent imaging studies using 3.0-T and 1.5-T scanners from GE, Siemens, and Philips. Four MRI-resistant adjustable shunt devices-proGAV2.0 (Miethke), Codman Certas Plus (Johnson & Johnson), Polaris (Sophysa), and Strata MR valve (Medtronic)-were fixed on the left temporal scalp. Routine MRI images, including T1-and T2-weighted imaging, fluid-attenuated inversion recovery, diffusion-weighted imaging (DWI), and magnetic resonance angiography (MRA), were obtained. We also compared artifacts between a 3.0-T scanner and a-1.5 T scanner in 4 patients.

RESULTS: The 3.0 T-scanners showed smaller artifacts than the 1.5-T scanners on DWI and MRA images for all shunt devices and scanners. In the other sequences, the results depended on the MRI scanner manufacturer; however, the GE 3.0-T scanner showed smaller artifacts in every sequence. This was also true in the 4 clinical cases.

CONCLUSIONS: A 3.0-T scanner is recommended over a 1.5-T scanner for patients with MRI-resistant adjustable shunt devices in the diagnosis of acute ischemic condition or when using GE scanners.

来源:World Neurosurg,查看原文。版权归原作者所有。

引用

  1. Amano Y, Kuroda N, Uchida D, et al. Unexpectedly Smaller Artifacts of 3.0-T Magnetic Resonance Imaging than 1.5 T: Recommendation of 3.0-T Scanners for Patients with Magnet-Resistant Adjustable Ventriculoperitoneal Shunt Devices. World Neurosurg. 2019. PMID: 31260847. doi:10.1016/j.wneu.2019.06.095

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。