利用不同锥形束CT曝光参数优化图像质量与辐射剂量

Optimization of image quality and radiation dose using different cone-beam CT exposure parameters.
PubMed 文献 · Eur J Radiol · 2019-04-22

研究对同一具尸体前臂在桡骨远端钢板植入前后共进行110次连续扫描,比较不同管电压与管电流-曝光时间乘积组合对图像质量和辐射剂量的影响;

要点

  • 厂商建议方案90 kV、5.0 mAs的DLP为111.91 mGy*cm、CTDIVOL为4.49 mGy,为最高辐射剂量。
  • 70 kV、2.0 mAs的DLP为20.34 mGy*cm、CTDIVOL为0.79 mGy,为剂量最低方案。
  • 较高管电压改善皮质骨、松质骨及整体图像质量的主观评分,改变mAs未见图像质量显著变化。

研究设计与样本

  • 对同一具尸体前臂在桡骨远端掌侧钢板植入前后进行110次连续扫描,设备为Carestream OnSight 3D Extremity System锥形束CT,采用不同管电压与管电流-曝光时间乘积组合,由五名读者独立盲法采用5点Likert量表进行主观图像质量评估。

关键影像表现/征象

  • 较高管电压改善皮质骨显示(p ≤ 0.038)、松质骨显示(p ≤ 0.001)及整体图像质量(p ≤ 0.027)。
  • 改变管电流-曝光时间乘积未显示图像质量的显著变化(p ≥ 0.063)。
  • 钢板植入后仅主观整体图像质量评分降低(p < 0.001)。
  • 不同扫描方案间客观图像质量分析无相关变化(SNR:p ≥ 0.125;CNR:p ≥ 0.086),但存在接骨板显著降低平均SNR和CNR(p < 0.001)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To investigate and optimize the impact of different exposure parameters on image quality and radiation dose for a latest generation orthopedic cone-beam CT system.

MATERIALS AND METHODS: 110 consecutive scans of the same cadaver forearm were performed before and after the insertion of a distal radius plate on the palmar radius to achieve highest intra-individual comparability. All scans were conducted on a latest generation cone-beam CT scanner (Carestream OnSight 3D Extremity System, Carestream Health, Rochester, NY, USA). Extremity imaging was performed using different combinations of tube voltage (kV) and tube current - exposure time product (mAs). Radiation dose (DLP and CTDIVOL) was recorded to widely varying combinations. Subjective and objective image quality analysis included a blinded evaluation by five different readers independently using 5-point-Likert scales.

RESULTS: Highest radiation dose was achieved using the manufacturers' suggested standard protocol (90-kV and 5.0 mAs with DLP of 111.91 mGy*cm and CTDIVOL of 4.49 mGy), while 70-kV and 2.0 mAs provided the most dose reduction with DLP of 20.34 mGy*cm and CTDIVOL of 0.79 mGy. Regarding subjective image quality, higher tube voltage improved depiction of cortical bone (p ≤ 0.038) and cancellous bone (p ≤ 0.001) as well as overall image quality (p ≤ 0.027). Changes of the tube current - exposure time product did not show significant alterations of image quality (p ≥ 0.063). After plate insertion, only the subjective overall image quality showed reduced subjective perception (p < 0.001). Between the different scan protocols, no relevant changes were observed in the objective image quality analysis (SNR: p ≥ 0.125; CNR: p ≥ 0.086). However, presence of osteosynthesis significantly lowered the mean SNR and CNR (p < 0.001).

CONCLUSION: Even with lowest exposure settings, orthopedic extremity CBCT revealed good overall image quality. The best result regarding subjective image quality was achieved with 85-kV / 4.7 mAs with a dose reduction of 18,9% compared to the manufacturer's recommended protocol (90-kV and 5.0 mAs).

来源:Eur J Radiol,查看原文。版权归原作者所有。