文献速递骨肌技术与安全CT

双层光谱计算机断层扫描用于金属伪影减少:小型与大型骨科器械对比

Dual-layer spectral computerized tomography for metal artifact reduction: small versus large orthopedic devices.
PubMed 文献 · Skeletal Radiol · 2019-06-01

这项前瞻性研究纳入12例金属骨科植入物患者,采用原型双层光谱探测器CT扫描,比较常规图像与40-200 keV虚拟单色图像。随能量水平升高伪影尺寸逐渐减小(p<0.01),200 keV时减小最明显,160-180 keV整体诊断图像质量最佳,小型植入物伪影减少优于大型硬件。

要点

  • 虚拟单色图像在160-180 keV整体诊断图像质量最佳,200 keV伪影尺寸减小最明显。
  • 随能量水平升高,骨-金属界面可见度(p=0.002)、条纹伪影程度(p=0.010)及距植入物1 cm处骨小梁显示(p=0.023)定性评分改善。
  • 小型植入物的伪影减少优于大型硬件。

核心结论

  • 双层光谱CT有助于减少金属伪影,虚拟单色图像诊断价值更高,尤其对较小植入物;160-180 keV的虚拟单能图像适合减少伪影。

研究设计与样本

  • 前瞻性研究,共12例患者,使用原型光谱探测器CT扫描金属骨科植入物,常规图像为120 kVp迭代重建,虚拟单色图像在40至200 keV之间以20 keV间隔生成,由两名骨肌放射科医生独立进行定性分析。

关键影像表现/征象

  • 随能量水平升高,伪影尺寸逐渐减小(p<0.01),常规图像与单色图像比较时200 keV减小最明显。
  • 定性评估显示160和180 keV水平整体诊断图像质量最佳。
  • 随能量水平升高,骨-金属界面可见度(p=0.002)、条纹伪影程度(p=0.010)及距植入物1 cm处骨小梁显示(p=0.023)的定性评分改善。
  • 距植入物5 cm处骨显示、软组织细节和整体诊断质量呈接近显著的改善趋势。
  • 亚组分析显示小型植入物伪影减少优于大型硬件。

临床应用与阅片要点

  • 虚拟单色图像在160-180 keV整体诊断图像质量最佳,可用于减少金属伪影。
  • 双层光谱CT减少金属伪影对较小植入物效果更优,大型硬件伪影减少相对有限。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

INTRODUCTION: Metal artifacts limit the diagnostic utility of computerized tomography (CT) for implant-related complications. Dual-layer spectral detector CT imaging has been suggested for artifact reduction. Our objective was to evaluate the utility of spectral CT in artifact reduction in patients with small and large metal implants.

METHODS: In this prospective study, patients with metallic orthopedic implants underwent CT imaging using a prototype spectral detector CT scanner. Conventional images were generated with iterative reconstruction at 120 kVp, and virtual monochromatic images were generated at 20-keV intervals between 40 to 200 keV. Conventional and monochromatic images were compared quantitatively using signal-to-noise ratio (SNR) and artifact improvement. Qualitative analysis was performed independently by two musculoskeletal radiologists and included six image quality indicators.

RESULTS: A total of 12 patients were scanned. In monochromatic images, as the energy level increased, the artifact size decreased progressively (p < 0.01). When conventional and monochromatic images were compared, maximum reduction was seen at 200 keV. Using qualitative assessments, 160 and 180 keV levels had the best overall diagnostic image quality. With increased energy level, there was improvement in qualitative ratings of bone-metal interface conspicuity (p = 0.002), degree of streak artifact (p = 0.010) and trabecular bone definition at 1 cm from implant (p = 0.023), and a trend towards significance for bone definition at 5 cm, soft tissue detail and overall diagnostic quality. Subgroup analysis revealed superior artifact reduction in small implants compared to large hardware.

DISCUSSION: Our results support the utility of dual-layer spectral CT in metal artifact reduction. Virtual monochromatic images were diagnostically superior, especially for smaller implants. Virtual monoenergetic images at 160-180 keV are ideal for reducing artifacts.

来源:Skeletal Radiol,查看原文。版权归原作者所有。