对比增强双能谱乳腺摄影(CESM)的诊断效能:一项纳入644个乳腺病变的回顾性研究

Diagnostic performance of contrast-enhanced dual-energy spectral mammography (CESM): a retrospective study involving 644 breast lesions.
PubMed 文献 · Radiol Med · 2019-06-27

本回顾性研究由五位放射科医生对644个乳腺病变(218个良性、426个恶性)分别用FFDM、FFDM+BUS和CESM评估,以病理或临床随访为金标准。CESM较FFDM和FFDM+BUS显著提高敏感度至93.2%、特异度至84.4%、阳性预测值至92.3%、阴性预测值至86.0%和准确度至90.2%,三种方法AUC比较有统计学差异(p<0.001)。CESM测量的肿瘤直径与病理测量一致性良好(r=0.891,p<0.0001),但倾向于高估尺寸(+7 mm)。

要点

  • CESM对乳腺病变诊断准确度高(90.2%),敏感度93.2%、特异度84.4%。
  • CESM较FFDM和FFDM+BUS在敏感度、特异度、PPV、NPV和准确度上均有提升。
  • CESM测量肿瘤大小与病理一致性良好(r=0.891),但平均高估约7 mm。

核心结论

  • CESM具有较高的诊断准确度,可被认为是评估乳腺病变的有用技术。

研究设计与样本

  • 回顾性研究,纳入644个乳腺病变(218个良性、426个恶性),由五位放射科医生分别评估FFDM、FFDM+BUS和CESM,以组织病理学或临床随访为金标准。

诊断标准与关键数值

  • CESM的敏感度为93.2%,较FFDM提高10.7%、较FFDM+BUS提高3.4%。
  • CESM的特异度为84.4%,较FFDM提高15.8%、较FFDM+BUS提高1.7%。
  • CESM的阳性预测值为92.3%,较FFDM提高26.8%、较FFDM+BUS提高3.6%。
  • CESM的阴性预测值为86.0%,较FFDM提高1.6%、较FFDM+BUS提高1.8%。
  • CESM的准确度为90.2%,较FFDM提高15.8%、较FFDM+BUS提高3.2%;三种方法AUC比较p<0.001。

临床应用与阅片要点

  • 作者认为CESM具有较高诊断准确度,可作为评估乳腺病变的有用技术。
  • CESM测量肿瘤直径与病理一致性良好(Spearman秩相关r=0.891,p<0.0001),但倾向于高估尺寸(+7 mm)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To evaluate the diagnostic performance of contrast-enhanced dual-energy spectral mammography (CESM) in comparison with that of full-field digital mammography (FFDM), either alone or accompanied with breast ultrasound (BUS) in a large series of patients/breast lesions (n = 644).

PATIENTS AND METHODS: In this retrospective study, five radiologists evaluated the lesions by three imaging modalities: FFDM, FFDM + BUS, and CESM and compared the imaging to the gold standard (histopathology or clinical follow-up). Diagnostic performance parameters and receiver operating characteristic (ROC) curves of CESM were calculated and compared to those of FFDM or FFDM + BUS (McNemar's test). Additionally, the reliability of tumor size measurement by CESM was compared with the histopathological measurement.

RESULTS: The study included 218 benign and 426 malignant lesions. 85% of benign and 93% of malignant lesions were adequately identified using CESM. With respect to FFDM and FFDM + BUS, CESM significantly increased sensitivity to 93.2% (+ 10.7% and + 3.4%, respectively); specificity to 84.4% (+ 15.8% and + 1.7%, respectively); PPV to 92.3% (+ 26.8% and + 3.6%, respectively); NPV to 86.0% (+ 1.6% and + 1.8%, respectively); and accuracy to 90.2% (+ 15.8% and + 3.2%, respectively). In the ROC curves analyses, the comparison among the three AUC values was also statistically significant (p < 0.001). Good agreement between tumor diameters measured using CESM and histopathology was observed (Spearman's rank correlation, r = 0.891, p < 0.0001), although this technique tended to produce an overestimation of the size (+ 7 mm).

CONCLUSIONS: CESM has high diagnostic accuracy and can be considered as a useful technique for the assessment of breast lesions.

来源:Radiol Med,查看原文。版权归原作者所有。