用于评估超声计算机断层成像声速定量准确性的油凝胶体模

Oil Gel-Based Phantom for Evaluating Quantitative Accuracy of Speed of Sound Measured in Ultrasound Computed Tomography.
PubMed 文献 · Ultrasound Med Biol · 2019-06-11

为评估乳腺超声计算机断层成像声速定量准确性,研制了含已知声速包涵体(水与盐水溶液)并可温控的油凝胶体模。油凝胶由SEBS 10% w/w加入石蜡油制成,可模拟脂肪乳腺组织到致密乳腺组织的声折射;43天内重建图像中油凝胶声速与衰减变化范围分别为0.7 m/s与0.03 dB/MHz/cm。盐水浓度在聚丙烯酰胺凝胶体模24小时内由1% w/w降至0.48% w/w,而油凝胶体模中保持恒定;作者结论认为该体模时间稳定性高,适合多中心分发并可能用于跨中心数据采集与图像重建标准化。

要点

  • 油凝胶体模可温控模拟脂肪乳腺组织到致密乳腺组织的声折射。
  • 43天内油凝胶重建图像声速与衰减变化范围分别为0.7 m/s与0.03 dB/MHz/cm。
  • 盐水浓度在聚丙烯酰胺凝胶体模24小时内由1% w/w降至0.48% w/w,油凝胶体模中恒定。

核心结论

  • 作者结论认为该油凝胶体模时间稳定性高,适合多中心分发,并可能用于跨中心数据采集与图像重建标准化。

研究设计与样本

  • 研制含已知声速包涵体(水或3.5% w/w氯化钠水溶液)并可温控的油凝胶体模,油凝胶由SEBS 10% w/w加入石蜡油制成,呈圆柱形并含可填充水或盐水溶液的棒状包涵体。

诊断标准与关键数值

  • 温度升高时水中声速升高而油凝胶中声速降低;温度<20.6°C时油凝胶声速高于水,温度>20.6°C时油凝胶声速低于水。
  • 43天内重建图像中油凝胶声速变化范围为0.7 m/s,衰减变化范围为0.03 dB/MHz/cm。
  • 聚丙烯酰胺凝胶体模中盐水浓度24小时内由1% w/w降至0.48% w/w,油凝胶体模中盐水浓度保持恒定。
  • 油凝胶体模的磁共振成像显示NiSO4溶液在体模中稳定保持42天。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

To evaluate the quantitative accuracy of the measured speed of sound in ultrasound computed tomography for breast imaging, it is necessary to use a phantom with inclusions whose speed of sound is known. Accordingly, a phantom with known-speed-of-sound inclusions (e.g., containing water and saltwater solution) under the control of temperature was developed. In addition, an oil gel was used as the phantom material for mimicking wave refraction from fatty breast tissue to dense breast tissue. The oil gel was generated by adding SEBS (styrene-ethylene/butylene-styrene, 10% w/w) to paraffin oil. The oil gel-based phantom has a cylindrical shape and contains rod-shaped inclusions that can be filled with water or saltwater solution (3.5% w/w sodium chloride in water). When temperature increases, the speed of sound in the water increases, while that in the oil gel decreases; in particular, the speed of sound in the oil gel was higher than that in the water at temperatures <20.6°C, while the speed of sound in the oil gel was lower than that in the water at temperatures >20.6°C. It has been reported that the speed of sound in dense breast tissue is higher than that in water, while that in fatty breast tissue is lower than that in water. Ultrasound is refracted owing to the difference between the speed of sound in the breast tissue and that in the background water. By controlling the temperatures of the oil gel and water, the oil gel-based phantom simulates the refraction of an ultrasound wave from fatty breast tissue to dense breast tissue. For 43 d, the variation ranges of the speed of sound and attenuation in the oil gel in the reconstructed images were 0.7 m/s and 0.03 dB/MHz/cm, respectively. The concentration of the saltwater solution in the polyacrylamide gel-based phantom decreased from 1% (w/w) to 0.48% (w/w) after 24 h, while that in the oil-gel-based phantom was constant. In addition, magnetic resonance imaging of the oil gel-based phantom revealed that NiSO4 solution was stably contained in the phantom for 42 d. It is therefore concluded that the liquid cannot penetrate the oil gel. This oil gel-based phantom with such high temporal stability is suitable for multicenter distribution and may be used for standardization of data acquisition and image reconstruction across centers.