超声剪切波弹性成像对≤4 cm实性肾实质肿块的诊断效能
研究纳入20名健康志愿者和117例患者,用剪切波弹性成像(SWE)鉴别≤4 cm实性肾实质肿块良恶性,并与常规超声比较。肾细胞癌的平均弹性、最小弹性和弹性比均显著低于良性病变(均p<0.001),SWE联合常规超声可将预测肾细胞癌的特异度提高至87.8%,但敏感度未提高。
要点
- 肾细胞癌的Emean、Emin和Eratio均显著低于良性病变(均p<0.001)。
- Emean、Emin和Eratio的诊断界值分别为9.15 kPa、3.55 kPa和1.99。
- SWE联合常规超声可将预测肾细胞癌的特异度提高至87.8%,敏感度未提高。
核心结论
- SWE参数Emean、Emin和Eratio在≤4 cm实性肾实质肿块中鉴别肾细胞癌与良性病变具有诊断价值,联合常规超声可提高特异度但不提高敏感度。
研究设计与样本
- 研究纳入20名健康志愿者和117例患者(68例肾细胞癌、49例良性病变),所有志愿者和患者均行常规超声和SWE检查,比较SWE与常规超声的诊断效能。
关键影像表现/征象
- 健康右肾和左肾皮质弹性值分别为4.7±1.7 kPa和4.5±1.5 kPa。
- 肾细胞癌Emean为7.2±2.5 kPa,低于良性病变的10.0±2.4 kPa(p<0.001)。
- 肾细胞癌Emin为2.5±2.4 kPa,低于良性病变的5.6±2.3 kPa(p<0.001)。
- 肾细胞癌Eratio为1.6±0.5,低于良性病变的2.2±0.6(p<0.001)。
诊断标准与关键数值
- Emean界值9.15 kPa的AUC为0.801。
- Emin界值3.55 kPa的AUC为0.832。
- Eratio界值1.99的AUC为0.806。
- Emean、Emin和Eratio联合常规超声预测肾细胞癌的特异度为87.8%,敏感度未提高。
临床应用与阅片要点
- 对≤4 cm实性肾实质肿块,可测量Emean、Emin和Eratio辅助鉴别肾细胞癌与良性病变。
- SWE联合常规超声有助于提高预测肾细胞癌的特异度,但敏感度未提高。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
The aim of the study was to analyze the diagnostic performance of shear wave elastography (SWE) in differentiating between malignant and benign solid renal parenchymal masses ≤4 cm, compared with conventional ultrasound. A total of 20 healthy volunteers and 117 patients had been included in this study. Conventional ultrasound and SWE were performed in all volunteers and patients. The elasticity of healthy cortex and the elastic parameters of tumors such as mean elasticity (Emean), minimum elasticity (Emin), maximum elasticity (Emax), standard deviation and elasticity ratio of the lesion to surrounding cortex (Eratio) were measured on SWE images. Diagnostic performance of SWE was compared with that of conventional ultrasound. The cortical elasticity values of healthy right and left kidneys were 4.7 ± 1.7 and 4.5 ± 1.5 kPa, respectively. Of the 117 renal tumors, 68 were renal cell carcinomas (RCCs) and 49 were benign. Emean, Emin and Eratio were significantly lower in RCCs compared with benign lesions: Emean 7.2 ± 2.5 kPa versus 10.0 ± 2.4 kPa, Emin 2.5 ± 2.4 kPa versus 5.6 ± 2.3 kPa, Eratio 1.6 ± 0.5 versus 2.2 ± 0.6 (all p values < 0.001). The cutoff values of 9.15 kPa for Emean, 3.55 kPa for Emin and 1.99 for Eratio had the highest areas under the receiver operating characteristics curve (0.801 for Emean, 0.832 for Emin and 0.806 for Eratio). Combining Emean, Emin and Eratio with conventional ultrasound improved the specificity for predicting RCCs to 87.8%, but the sensitivity was not increased.