基于MRI质子密度脂肪分数的脾脏脂肪测量源于噪声重建

Measurement of spleen fat on MRI-proton density fat fraction arises from reconstruction of noise.
PubMed 文献 · Abdom Radiol (NY) · 2019-01-01

本研究为前瞻性辅助研究,在47例已知或疑似非酒精性脂肪性肝病的成人和儿童中于3T比较了脾脏质子密度脂肪分数(PDFF)的CSE-MRI测量与MRS。四种CSE-MRI方法测得的脾脏PDFF均显著高于MRS(中位数2.3、1.9、2.2、2.0% vs 1.0%,p<0.0001),且CSE-MRI方法间相关性弱且多重比较校正后均不显著。作者结论认为人脾脏中非零PDFF值为伪影。

要点

  • 脾脏CSE-MRI测得的PDFF中位数为2.0–2.3%,而MRS为1.0%。
  • 四种CSE-MRI方法测得的脾脏PDFF均显著高于MRS(p<0.0001)。
  • 不同CSE-MRI方法间脾脏PDFF相关性弱,校正后均不显著。

核心结论

  • CSE-MRI较MRS高估脾脏PDFF且相关CSE-MRI方法间一致性差,提示人脾脏中非零PDFF值为伪影。

研究设计与样本

  • 前瞻性设计的辅助研究,纳入47例已知或疑似非酒精性脂肪性肝病的成人和儿童(20例女性、27例男性;12例成人中位55岁,35例儿童中位12岁),在3T行MRI-M、MRI-C、Hi-SNR MRI-M、Hi-SNR MRI-C及MRS估计脾脏PDFF。

诊断标准与关键数值

  • 脾脏PDFF中位数:MRS为1.0%,MRI-M为2.3%,Hi-SNR MRI-M为1.9%,MRI-C为2.2%,Hi-SNR MRI-C为2.0%。
  • 四种CSE-MRI方法测得的脾脏PDFF均显著高于MRS,p<0.0001。
  • 不同CSE-MRI方法间脾脏PDFF的Spearman秩相关最高为MRI-M与Hi-SNR MRI-M之间的0.295,多重比较校正后均不显著。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: This study compares splenic proton density fat fraction (PDFF) measured using confounder-corrected chemical shift-encoded (CSE)-MRI to magnetic resonance spectroscopy (MRS) in human patients at 3T.

METHODS: This was a prospectively designed ancillary study to various previously described single-center studies performed in adults and children with known or suspected nonalcoholic fatty liver disease. Patients underwent magnitude-based MRI (MRI-M), complex-based MRI (MRI-C), high signal-to-noise variants (Hi-SNR MRI-M and Hi-SNR MRI-C), and MRS at 3T for spleen PDFF estimation. PDFF from CSE-MRI methods were compared to MRS-PDFF using Wilcoxon signed-rank tests. Demographics were summarized descriptively. Spearman's rank correlations were computed pairwise between CSE-MRI methods. Individual patient measurements were plotted for qualitative assessment. A significance level of 0.05 was used.

RESULTS: Forty-seven patients (20 female, 27 male) including 12 adults (median 55 years old) and 35 children (median 12 years old). Median PDFF estimated by MRS, MRI-M, Hi-SNR MRI-M, MRI-C, and Hi-SNR MRI-C was 1.0, 2.3, 1.9, 2.2, and 2.0%. The four CSE-MRI methods estimated statistically significant higher spleen PDFF values compared to MRS (p < 0.0001 for all). Pairwise associations in spleen PDFF values measured by different CSE-MRI methods were weak, with the highest Spearman's rank correlations being 0.295 between MRI-M and Hi-SNR MRI-M; none were significant after correction for multiple comparisons. No qualitative relationship was observed between PDFF measurements among the various methods.

CONCLUSION: Overestimation of PDFF by CSE-MRI compared to MRS and poor agreement between related CSE-MRI methods suggest that non-zero PDFF values in human spleen are artifactual.