基于R2弛豫测量法的MRI评估肝脏铁含量:两种方法的比较
研究比较了本院自建分析与FDA批准的商用第三方软件(FerriScan®)在1.5T MRI R2弛豫测量法评估肝脏铁浓度(LIC)中的可重复性与准确性。共纳入175例患者的474次检查,R2值范围31.5-334.8 s-1,两种方法相关系数r=0.87(p<0.001),Bland-Altman平均偏差+21.5 s-1;FerriScan®报告的LIC与本院基于其R2估算的LIC高度一致(r=1.0,p<0.001)。结论认为两种方法在LIC评估中具有可重复性。
要点
- 本院自建分析与FerriScan®商用软件测得的R2值高度相关(r=0.87,p<0.001)。
- FerriScan®报告的LIC与本院基于其R2估算的LIC几乎完全一致(r=1.0,p<0.001)。
核心结论
- R2弛豫测量法MRI评估肝脏铁浓度时,FDA批准的商用软件与本院自建分析方法之间具有可重复性。
研究设计与样本
- 比较研究,纳入175例患者的474次MR检查,平均年龄10.4±4.2岁(范围1-18岁),其中女童254次、男童220次;使用1.5T扫描仪,多回波自旋回波序列固定TR、TE值分别为6 ms、9 ms、12 ms、15 ms、18 ms(间隔3 ms),于中间层面勾画覆盖全肝的感兴趣区计算平均弛豫率R2;以FerriScan®(Resonance Health,澳大利亚)商用软件结果为外部参照,采用相关系数与Bland-Altman差异图评估本院自建分析与外部公司报告值之间的一致性。
诊断标准与关键数值
- 肝脏铁浓度(LIC)范围为0.6至43 mg/g干组织,覆盖正常至极高铁水平;R2值范围为31.5至334.8 s-1;本院自建分析与商用软件R2值的相关系数r=0.87(p<0.001);Bland-Altman R2差异图平均偏差+21.5 s-1(两倍标准差范围-47.0至+90.0 s-1);FerriScan®报告的LIC与本院基于其R2估算的LIC相关系数r=1.0(p<0.001)。
临床应用与阅片要点
- 作者结论认为,R2弛豫测量法MRI评估肝脏铁浓度在FDA批准的商用软件与本院自建分析方法之间具有可重复性;未给出具体阅片或报告操作建议。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To compare the reproducibility and accuracy of R2-relaxometry MRI for estimation of liver iron concentration (LIC) between in-house analysis and FDA-approved commercially available third party results.
METHODS: All MR studies were performed on a 1.5T scanner. Multi-echo spin-echo scans with a fixed TR and increasing TE values of 6 ms, 9 ms, 12 ms, 15 ms, and 18 ms (spaced at 3 ms intervals) were used. Post-processing of the images to calculate mean relaxivity, R2, included drawing of regions of interest to include the whole liver on mid-slice. The relationship between liver R2 values and estimated LIC calculated with in-house analysis and values reported by an external company (FerriScan®, Resonance Health, Australia) were assessed with correlation coefficients and Bland-Altman difference plots. Continuous variables are presented as mean ± standard deviation. Significance was set at p value < 0.05.
RESULTS: 474 studies from 175 patients were included in the study (mean age 10.4 ± 4.2 years (range 1-18 years); 254 studies from girls, 220 studies from boys). LIC ranged from 0.6 to 43 mg/g dry tissue, covering a broad range from normal levels to extremely high iron levels. Linearity between proprietary and in-house methods was excellent across the observed range for R2 (31.5 to 334.8 s-1); showing a correlation coefficient of r = 0.87, p < 0.001. Bland-Altman R2 difference plot between the two methods shows a mean bias of + 21.5 s-1 (range - 47.0 to + 90.0 s-1 between two standard deviations). LIC reported by FerriScan® compared with LIC estimated in-house with R2 as reported by FerriScan® agreed strongly, (r = 1.0, p < 0.001).
CONCLUSION: R2 relaxometry MR imaging for liver iron concentration estimation is reproducible between proprietary FDA-approved commercial software and in-house analysis methods.