急诊科无意识患者超快速扫描模式与常规扫描模式CT图像运动伪影的比较

Comparison of Motion Artifacts on CT Images Obtained in the Ultrafast Scan Mode and Conventional Scan Mode for Unconscious Patients in the Emergency Department.
PubMed 文献 · AJR Am J Roentgenol · 2019-06-05

回顾性评估60例急诊无意识患者的CT图像,30例使用超快速扫描模式的新一代CT扫描仪,30例使用常规CT扫描仪。两位放射科医师独立评估主动脉、肺、膈肌、肝和肾的运动伪影,超快速扫描模式在主动脉根部、下肺、膈肌、肝和肾的运动伪影显著减少,但在主动脉弓、胸降主动脉、腹主动脉和上肺无显著差异。

要点

  • 超快速扫描模式可显著减少主动脉根部、下肺、膈肌、肝和肾的运动伪影
  • 主动脉弓、胸降主动脉、腹主动脉和上肺的运动伪影两组无显著差异
  • 两位放射科医师对运动伪影评分的一致性良好(κ=0.80-0.93)

核心结论

  • 超快速扫描模式CT图像在急诊无意识患者评估中运动伪影显著减少,该技术有望用于急诊科诊断性CT。

研究设计与样本

  • 回顾性研究,共60例急诊无意识患者,30例使用超快速扫描模式的新一代CT扫描仪,30例使用常规CT扫描仪,两位放射科医师独立评估运动伪影。

关键影像表现/征象

  • 运动伪影按严重程度以4分制分级评估。
  • 超快速扫描模式组主动脉根部运动伪影较常规扫描组显著减少(p=0.0003)。
  • 超快速扫描模式组下肺运动伪影较常规扫描组显著减少(p=0.011)。
  • 超快速扫描模式组膈肌运动伪影较常规扫描组显著减少(p=0.0047)。
  • 超快速扫描模式组肝运动伪影较常规扫描组显著减少(p=0.0026),肾运动伪影亦显著减少(p=0.019)。

诊断标准与关键数值

  • 运动伪影严重程度采用4分制分级评估。
  • 两位放射科医师对运动伪影评分的一致性良好(κ=0.80-0.93)。
  • 主动脉弓、胸降主动脉、腹主动脉和上肺的运动伪影两组间无显著差异。

临床应用与阅片要点

  • 原文结论认为超快速扫描模式CT图像运动伪影显著减少,该技术有望用于急诊科诊断性CT。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE. The purpose of this study was to investigate whether, in the evaluation of unconscious patients in the emergency department, a new-generation CT scanner that acquires images in ultrafast scan mode (large coverage, fast rotation, high helical pitch) would reduce motion artifacts on whole-body CT images in comparison with those on images obtained with a conventional CT scanner. MATERIALS AND METHODS. Images of a total of 60 unconscious patients presenting to the emergency department were evaluated retrospectively. Of the 60 patients, 30 underwent CT with a new-generation scanner that acquires images in the ultrafast mode, and 30 underwent CT with a conventional scanner. Two radiologists independently evaluated motion artifacts in the aorta, lung, diaphragm, liver, and kidneys. The motion artifacts were graded in severity on a 4-point scale. A value of p < 0.05 was considered to indicate a statistically significant difference. RESULTS. Interobserver agreement on motion artifact ratings was good (κ = 0.80-0.93). Images obtained with the new-generation CT scanner showed a significant reduction in motion artifacts in the aortic root (p = 0.0003), lower lungs (p = 0.011), diaphragm (p = 0.0047), liver (p = 0.0026), and kidneys (p = 0.019). However, there were no significant differences between the two groups with respect to motion artifacts of the aortic arch, thoracic descending aorta, abdominal aorta, and upper lungs. CONCLUSION. CT images obtained in the ultrafast scan mode in the evaluation of unconscious patients had a significant reduction in motion artifacts. The ultrafast technique is expected to be useful for diagnostic CT in the emergency department.