PROPELLER技术在3T眼眶MRI中使用眼表面线圈减少眼球运动伪影并提高图像质量的效能
研究纳入46例使用眼表面线圈行3T眼眶MRI的患者,21例行T2W和T1 FLAIR PROPELLER序列,25例行常规T2W和T1W序列;两名放射科医师独立采用五分法评估。PROPELLER序列在所有病例中评分均高于常规序列(p<0.05),运动伪影更少、图像质量更高、解剖结构和病变显示更清晰。T2W PROPELLER在晶状体、玻璃体、泪腺和病变中SNR高于常规T2W(p<0.05),而T1 FLAIR PROPELLER在晶状体、玻璃体、内直肌、外直肌、颞肌和视神经后半部SNR低于常规T1W(p<0.05)。
要点
- PROPELLER序列在减少眼球运动伪影和提高眼眶MRI图像质量方面评分均高于常规序列(p<0.05)。
- T2W PROPELLER在晶状体、玻璃体、泪腺和病变的SNR高于常规T2W(p<0.05)。
- T1 FLAIR PROPELLER在晶状体、玻璃体、内直肌、外直肌、颞肌和视神经后半部SNR低于常规T1W(p<0.05)。
核心结论
- PROPELLER可有效减少眼球运动伪影并提高图像质量,在使用眼表面线圈的3T眼眶MRI中对清晰显示解剖结构和病变具有重要作用。
研究设计与样本
- 前瞻性对照研究,共46例患者行3T眼眶MRI并使用眼表面线圈:21例行增强前轴位T2W PROPELLER和T1 FLAIR PROPELLER成像,25例行常规T2W和T1W成像;由两名放射科医师独立采用五分法评估,并测量解剖结构和病变的SNR。
关键影像表现/征象
- PROPELLER图像较非PROPELLER图像显示更少的运动伪影和更高的图像质量。
- PROPELLER图像对解剖结构和病变的显示较非PROPELLER图像更清晰。
诊断标准与关键数值
- 采用五分法评估眼球运动伪影、解剖结构显示、病变显示和图像质量,两名医师间一致性良好至优秀(kappa值0.79至0.91)。
- PROPELLER序列在所有病例中的评分均高于常规序列(p<0.05)。
- T2W PROPELLER在晶状体、玻璃体、泪腺和病变中的SNR高于常规T2W(p<0.05)。
- T1 FLAIR PROPELLER在晶状体、玻璃体、内直肌、外直肌、颞肌和视神经后半部中的SNR低于常规T1W(p<0.05)。
临床应用与阅片要点
- 在使用眼表面线圈行3T眼眶MRI时,PROPELLER序列可减少眼球运动伪影并提高图像质量。
- T2W PROPELLER可提高晶状体、玻璃体、泪腺和病变的SNR,而T1 FLAIR PROPELLER在上述部分结构的SNR低于常规T1W。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To evaluate the efficacy of periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) in reducing ocular motion artefacts and improving image quality of orbital magnetic resonance imaging (MRI) at 3 T using an eye surface coil.
MATERIALS AND METHODS: Forty-six patients underwent orbital 3 T MRI using an eye surface coil. The pre-contrast axial T2-wighted (W) PROPELLER and T1 fluid-attenuated inversion recovery (FLAIR) PROPELLER imaging were performed on 21 patients, and conventional T2W and T1W imaging were performed on 25 patients. The ocular motion artefacts, delineation of anatomical structures, depiction of lesions, and image quality were evaluated independently by two radiologists using a five-point scale. The signal-to-noise ratios (SNRs) of anatomical structures and lesions were measured.
RESULTS: The interobserver agreement was good to excellent (kappa values from 0.79 to 0.91). PROPELLER sequences had higher scoring than the conventional sequences in all cases (p<0.05). PROPELLER images showed fewer motion artefacts, higher image quality, and more clear delineation of anatomical structures and lesions than the non-PROPELLER images. T2W PROPELLER images produced higher SNRs in lens, vitreous body, lacrimal gland, and lesions than conventional T2W images (p<0.05). T1 FLAIR PROPELLER images showed lower SNRs in lens, vitreous body, medial rectus, lateral rectus, temporalis, and posterior half of optic nerve than conventional T1W images (p<0.05).
CONCLUSION: PROPELLER can effectively reduce ocular motion artefacts and improve image quality, which plays an important role in clearly delineating anatomical structures and lesions on orbital 3 T MRI using an eye surface coil.
引用
- Jiang H, Wang S, Xian J, et al. Efficacy of PROPELLER in reducing ocular motion artefacts and improving image quality of orbital MRI at 3 T using an eye surface coil. Clin Radiol. 2019. PMID: 31221469. doi:10.1016/j.crad.2019.05.014
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。