腹主动脉与下腔静脉的定量时谐超声弹性成像研究

Quantitative Time-Harmonic Ultrasound Elastography of the Abdominal Aorta and Inferior Vena Cava.
PubMed 文献 · Ultrasound Med Biol · 2019-06-11

本研究对20名健康志愿者在口服1升水及被动抬腿前后行下腔静脉与腹主动脉时谐超声弹性成像,测量剪切波速度以评估血管壁硬度。下腔静脉剪切波速度随容量负荷增加而升高(从1.71±0.1 m/s升至1.95±0.1 m/s,p<0.001),腹主动脉剪切波速度在水合后无显著变化(2.14±0.13 m/s vs. 2.15±0.16 m/s,p=0.792)。时谐超声弹性成像可量化腹部大血管壁硬度,且对下腔静脉不同容量和压力状态敏感。

要点

  • 下腔静脉剪切波速度随容量负荷增加而升高,对容量状态敏感。
  • 腹主动脉剪切波速度在水合后无显著变化。
  • 腹主动脉剪切波速度在所有实验中均显著高于下腔静脉(p<0.001)。

核心结论

  • 时谐超声弹性成像能够量化腹部大血管壁硬度,并且对下腔静脉的不同容量和压力状态敏感。

研究设计与样本

  • 在20名健康志愿者中,于口服1升水前后以及被动抬腿前或被动抬腿期间行下腔静脉和腹主动脉时谐超声弹性成像,从多频谐波波场生成全视野剪切波速度复合图,并记录血压、每搏输出量、心输出量和脉搏波速度。

关键影像表现/征象

  • 下腔静脉剪切波速度在口服水前为1.71 ± 0.1 m/s,被动抬腿期间为1.82 ± 0.1 m/s,水合后为1.87 ± 0.1 m/s,水合加被动抬腿后为1.95 ± 0.1 m/s(p < 0.001)。
  • 腹主动脉剪切波速度在水合后无显著变化(2.14 ± 0.13 m/s vs. 2.15 ± 0.16 m/s;p = 0.792)。
  • 腹主动脉剪切波速度在所有实验中均显著高于下腔静脉(p < 0.001)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

The purpose of this study was to evaluate the sensitivity of quantitative time-harmonic ultrasound elastography (THE) of the inferior vena cava (IVC) and abdominal aorta (AA) to changes in central volume status. THE of the IVC and AA was performed in 20 healthy volunteers before and after oral intake of 1 L of water and before or during passive leg raising to augment venous filling. Compound maps of shear wave speed (SWS) as surrogate measures of vessel wall stiffness were generated within the full field of view from multifrequency harmonic wave fields. SWS was measured in regions of the IVC and AA. Blood pressure, stroke volume, cardiac output and pulse wave velocity were recorded. Statistical significance of SWS changes was tested using one-way repeated-measures analysis of variance. SWS measured in the IVC increased from 1.71 ± 0.1 m/s before water intake to 1.82 ± 0.1 m/s during passive leg raising and, further, to 1.87 ± 0.1 m/s after hydration and to 1.95 ± 0.1 m/s with hydration plus passive leg raising (p < 0.001). SWS in the AA did not change significantly after hydration (2.14 ± 0.13 m/s vs. 2.15 ± 0.16 m/s; p = 0.792). SWS was significantly higher in the AA than in the IVC across all experiments (p < 0.001). Water drinking did not significantly influence blood pressure, pulse wave velocity and cardiac output (all p values >0.1), whereas stroke volume increased significantly (p = 0.031). Time-harmonic ultrasound elastography enables quantification of the wall stiffness of the large abdominal vessels and is sensitive to different volume and pressure states in the IVC.