人脑MR指纹弛豫测量法的重现性与重复性研究
这项前瞻性研究于2017年7月至2018年1月在4个中心使用同一制造商10台不同型号扫描仪,对10名健康志愿者行二维稳态自由进动MR指纹成像并生成T1、T2图,评估跨扫描仪重现性与扫描仪内重复性;T1和T2的相对偏差在不同脑区为1.8%–20.9%,实性组织高于脑脊液,T1较T2更稳定;跨扫描仪平均T1、T2相对偏差的置信区间半宽分别为3.4%和8.0%,扫描仪内重复性半宽T1为2.0%–3.1%、T2为3.1%–7.9%。
要点
- T1和T2在实性脑组织中的重现性与重复性均高于脑脊液,且T1较T2更稳定。
- 跨扫描仪平均T1相对偏差95%CI半宽3.4%,平均T2为8.0%。
- 扫描仪内重复性相对偏差95%CI半宽T1为2.0%–3.1%,T2为3.1%–7.9%。
核心结论
- 本研究提供了健康志愿者脑组织稳态自由进动MR指纹T1、T2弛豫测量法的重现性与重复性数据,实性脑区高于脑脊液,T1高于T2。
研究设计与样本
- 前瞻性研究,2017年7月至2018年1月在4个中心使用同一制造商10台不同型号扫描仪,纳入10名健康志愿者(平均年龄28.5岁±6.9,男8例、女2例),采用基于稳态自由进动序列的二维层面选择性MR指纹成像并做外部射频发射场(B1+)校正生成T1、T2图。
诊断标准与关键数值
- T1和T2在脑内不同区域的相对偏差(95%CI)范围为1.8%–20.9%,实性组织高于脑脊液。
- 跨扫描仪平均T1相对偏差的95%CI半宽为3.4%,平均T2为8.0%。
- 扫描仪内重复性相对偏差的95%CI半宽:T1为2.0%–3.1%,T2为3.1%–7.9%。
临床应用与阅片要点
- 解读脑MR指纹T1、T2定量值时,需考虑脑区差异:实性组织重现性优于脑脊液,T1优于T2。
- 跨扫描仪比较T1、T2时,可参考本研究报告的相对偏差95%CI半宽(T1 3.4%、T2 8.0%)评估测量波动。
- 同一扫描仪重复测量时,可参考扫描仪内重复性半宽(T1 2.0%–3.1%、T2 3.1%–7.9%)判断变化是否超出测量误差。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Background Only sparse literature investigates the reproducibility and repeatability of relaxometry methods in MRI. However, statistical data on reproducibility and repeatability of any quantitative method is essential for clinical application. Purpose To evaluate the reproducibility and repeatability of two-dimensional fast imaging with steady-state free precession MR fingerprinting in vivo in human brains. Materials and Methods Two-dimensional section-selective MR fingerprinting based on a steady-state free precession sequence with an external radiofrequency transmit field, or B1+, correction was used to generate T1 and T2 maps. This prospective study was conducted between July 2017 and January 2018 with 10 scanners from a single manufacturer, including different models, at four different sites. T1 and T2 relaxation times and their variation across scanners (reproducibility) as well as across repetitions on a scanner (repeatability) were analyzed. The relative deviations of T1 and T2 to the average (95% confidence interval) were calculated for several brain compartments. Results Ten healthy volunteers (mean age ± standard deviation, 28.5 years ± 6.9; eight men, two women) participated in this study. Reproducibility and repeatability of T1 and T2 measures in the human brain varied across brain compartments (1.8%-20.9%) and were higher in solid tissues than in the cerebrospinal fluid. T1 measures in solid tissue brain compartments were more stable compared with T2 measures. The half-widths of the confidence intervals for relative deviations were 3.4% for mean T1 and 8.0% for mean T2 values across scanners. Intrascanner repeatability half-widths of the confidence intervals for relative deviations were in the range of 2.0%-3.1% for T1 and 3.1%-7.9% for T2. Conclusion This study provides values on reproducibility and repeatability of T1 and T2 relaxometry measured with fast imaging with steady-state free precession MR fingerprinting in brain tissues of healthy volunteers. Reproducibility and repeatability are considerably higher in solid brain compartments than in cerebrospinal fluid and are higher for T1 than for T2. © RSNA, 2019 Online supplemental material is available for this article. See also the editorial by Barkhof and Parker in this issue.