基于可定制可塑形聚乙烯醇冷冻凝胶材料与3D打印的超声成像医学体模的开发与表征:用于婴儿高频颅脑超声
本研究提出一种经济且易定制的体模制造方法,以聚乙烯醇冷冻凝胶模拟软组织、3D打印聚合物模拟硬组织,并表征材料声学特性随添加剂浓度的变化;该方法被用于制作婴儿高频颅脑超声体模,可再现经前囟新生儿超声图像的主要特征,深度达8 mm。
要点
- 聚乙烯醇冷冻凝胶配方在多次冻融循环中稳定,可制作性能可控的多层软组织模拟层。
- 材料声速、衰减系数和机械阻抗随丙二醇与氧化铝颗粒浓度变化而被表征。
- 所制婴儿颅脑超声体模可再现经前囟新生儿超声图像主要特征,深度达8 mm。
核心结论
- 提出一种基于聚乙烯醇冷冻凝胶与3D打印聚合物的经济、易定制体模制造方法,并成功应用于婴儿高频颅脑超声体模,可再现经前囟新生儿超声图像主要特征至8 mm深度。
研究设计与样本
- 体模制造方法学研究:以聚乙烯醇基冷冻凝胶模拟软组织、3D打印聚合物模拟硬组织及模塑支撑元件,表征材料声学特性随添加剂浓度变化,并制作婴儿高频颅脑超声体模验证。
关键影像表现/征象
- 聚乙烯醇冷冻凝胶可形成薄而耐用的软组织模拟层及多层结构。
- 材料声学特性(声速、衰减系数、机械阻抗)随丙二醇和氧化铝颗粒浓度变化而被表征。
- 所提聚乙烯醇配方在多次冻融循环中稳定,可制造性能可控的多层材料。
- 婴儿颅脑超声体模可再现经前囟新生儿超声图像的主要特征,深度达8 mm。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
This work presents an affordable and easily customizable methodology for phantom manufacturing, which can be used to mimic different anatomic organs and structures. This methodology is based on the use of polyvinyl alcohol-based cryogels as a physical substitute for biologic soft tissues and of 3-D printed polymers for hard tissues, moulding and supporting elements. Thin and durable soft-tissue mimicking layers and multilayer arrangements can be obtained using these materials. Special attention was paid to the acoustic properties (sound speed, attenuation coefficient and mechanical impedance) of the materials developed to simulate soft tissues. These properties were characterized as a function of the additives concentration (propylene-glycol and alumina particles). The polyvinyl alcohol formulation proposed in this work is stable over several freeze-thaw cycles, allowing the manufacturing of multilayer materials with controlled properties. The manufacturing methodology presented was applied to the development of a phantom for high-frequency cranial ultrasonography in infants. This phantom was able to reproduce the main characteristics of the ultrasound images obtained in neonates through the anterior fontanel, down to 8-mm depth.
引用
- Elvira L, Durán C, Higuti RT, et al. Development and Characterization of Medical Phantoms for Ultrasound Imaging Based on Customizable and Mouldable Polyvinyl Alcohol Cryogel-Based Materials and 3-D Printing: Application to High-Frequency Cranial Ultrasonography in Infants. Ultrasound Med Biol. 2019. PMID: 31128769. doi:10.1016/j.ultrasmedbio.2019.04.030
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。