颈动脉分叉处自适应速度复合血流速度矢量成像的体内研究

In Vivo Blood Velocity Vector Imaging Using Adaptive Velocity Compounding in the Carotid Artery Bifurcation.
PubMed 文献 · Ultrasound Med Biol · 2019-05-09

该研究提出一种自适应速度复合技术,利用双角度平面波采集的斑点追踪轴向位移估计二维速度矢量场。直血管实验显示最大速度幅值偏差为-3.7%(标准差2.8%),角度偏差为-0.16°(标准差0.41°)。在体可视化了两条健康颈动脉和三条病变颈动脉的复杂血流模式,矢量复杂度随管壁不规则程度增加而升高。

要点

  • 自适应速度复合技术可基于双角度平面波斑点追踪估计颈动脉二维速度矢量场。
  • 直血管实验中速度幅值偏差为-3.7%(标准差2.8%),角度偏差为-0.16°(标准差0.41°)。
  • 矢量复杂度随颈动脉管壁不规则程度增加而升高,可能有助于动脉粥样硬化早期检测。

核心结论

  • 自适应速度复合技术能在体内可视化并量化颈动脉分叉处复杂血流模式,矢量复杂度可能是有助于动脉粥样硬化早期检测的临床参数。

研究设计与样本

  • 方法学研究:使用7.8 MHz线阵探头连接Verasonics Vantage超声系统,进行直血管实验及两条健康与三条病变颈动脉的体内成像,以斑点追踪从双角度平面波采集估计轴向位移并计算二维速度矢量场。

关键影像表现/征象

  • 体内可视化了两条健康颈动脉和三条病变颈动脉的复杂血流模式。
  • 矢量复杂度随管壁不规则程度增加而升高。

诊断标准与关键数值

  • 直血管实验中最大速度幅值偏差为-3.7%(标准差2.8%)。
  • 直血管实验中角度偏差为-0.16°(标准差0.41°)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Visualization and quantification of blood flow are considered important for early detection of atherosclerosis and patient-specific diagnosis and intervention. As conventional Doppler imaging is limited to 1-D velocity estimates, 2-D and 3-D techniques are being developed. We introduce an adaptive velocity compounding technique that estimates the 2-D velocity vector field using predominantly axial displacements estimated by speckle tracking from dual-angle plane wave acquisitions. Straight-vessel experiments with a 7.8-MHz linear array transducer connected to a Verasonics Vantage ultrasound system revealed that the technique performed with a maximum velocity magnitude bias and angle bias of -3.7% (2.8% standard deviation) and -0.16° (0.41° standard deviation), respectively. In vivo, complex flow patterns were visualized in two healthy and three diseased carotid arteries and quantified using a vector complexity measure that increased with increasing wall irregularity. This measure could potentially be a relevant clinical parameter which might aid in early detection of atherosclerosis.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。