文献速递骨肌MR

AcidoCEST-UTE MRI 用于评估 3 T 下关节组织细胞外 pH 的可行性研究

AcidoCEST-UTE MRI for the Assessment of Extracellular pH of Joint Tissues at 3 T.
PubMed 文献 · Invest Radiol · 2019-01-01

本研究在 3 T 下用 acidoCEST-UTE MRI 序列,在液体体模和含软骨、半月板的组织体模中评估了碘普罗胺、碘帕醇和碘海醇的磁化传递率不对称、射频功率失配和相对饱和传递,结果显示碘帕醇和碘海醇最有前景,其中碘海醇的射频功率失配拟合在软骨、半月板和液体间相关系数最强,结论为利用 acidoCEST-UTE MRI 测量软骨和半月板细胞外 pH 是可行的。

要点

  • acidoCEST-UTE MRI 可在 3 T 下测量软骨和半月板的细胞外 pH,技术可行。
  • 碘帕醇和碘海醇在 pH 6.2–7.8 范围内表现浓度非依赖,且可在有或无缓冲液及生理温度下使用。
  • 碘海醇的射频功率失配拟合在软骨、半月板和液体间相关系数最强,其比值在液体与不同组织类型间相似。

核心结论

  • 利用 acidoCEST-UTE MRI 测量软骨和半月板的细胞外 pH 是可行的。

研究设计与样本

  • 可行性研究,在 3 T 下用 acidoCEST-UTE MRI 序列成像液体体模(碘普罗胺、碘帕醇、碘海醇,pH 6.2–7.8,不同浓度、温度及缓冲液浓度)和含软骨与半月板的组织体模(碘帕醇、碘海醇),计算相关系数和变异系数,并用配对 Wilcoxon 秩和检验评估统计学显著差异。

诊断标准与关键数值

  • 射频功率失配和相对饱和传递分析显示碘帕醇与碘海醇最有前景。
  • 碘帕醇和碘海醇在 pH 6.2–7.8 范围内表现为浓度非依赖,且可在有或无缓冲液及生理温度下使用。
  • 碘海醇的射频功率失配拟合在软骨、半月板和液体间相关系数最强。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The goal of this study was to demonstrate feasibility of measuring extracellular pH in cartilage and meniscus using acidoCEST technique with a 3-dimensional ultrashort echo time readout (acidoCEST-UTE) magnetic resonance imaging (MRI).

MATERIALS AND METHODS: Magnetization transfer ratio asymmetry, radiofrequency (RF) power mismatch, and relative saturation transfer were evaluated in liquid phantoms for iopromide, iopamidol, and iohexol over a pH range of 6.2 to 7.8, at various agent concentrations, temperatures, and buffer concentrations. Tissue phantoms containing cartilage and meniscus were evaluated with the same considerations for iopamidol and iohexol. Phantoms were imaged with the acidoCEST-UTE MRI sequence at 3 T. Correlation coefficients and coefficients of variations were calculated. Paired Wilcoxon rank-sum tests were used to evaluate for statistically significant differences.

RESULTS: The RF power mismatch and relative saturation transfer analyses of liquid phantoms showed iopamidol and iohexol to be the most promising agents for this study. Both these agents appeared to be concentration independent and feasible for use with or without buffer and at physiologic temperature over a pH range of 6.2 to 7.8. Ultimately, RF power mismatch fitting of iohexol showed the strongest correlation coefficients between cartilage, meniscus, and fluid. In addition, ratiometric values for iohexol are similar among liquid as well as different tissue types.

CONCLUSIONS: Measuring extracellular pH in cartilage and meniscus using acidoCEST-UTE MRI is feasible.