控制预压条件下脂肪与病灶应变比和腺体与病灶应变比在超声鉴别乳腺不确定及可疑病灶中的比较

Comparison of the Fat-to-Lesion Strain Ratio and the Gland-to-Lesion Strain Ratio With Controlled Precompression in Characterizing Indeterminate and Suspicious Breast Lesions on Ultrasound Imaging.
PubMed 文献 · J Ultrasound Med · 2019-05-29

本研究对39例连续患者的53个乳腺病灶在控制预压技术下行超声弹性成像,比较脂肪与病灶应变比(FLR)和腺体与病灶应变比(GLR)的诊断性能及观察者间可重复性。恶性病灶的FLR和GLR均显著高于良性病灶(P < .05);FLR的准确率和特异度高于GLR(准确率79.2%对60.4%,特异度87.5%对50.0%)。预压控制在小于25%时,FLR和GLR测量的观察者间一致性均优(ICC分别为0.853和0.779)。

要点

  • 恶性乳腺病灶的FLR和GLR均显著高于良性病灶,差异有统计学意义(P < .05)。
  • FLR诊断乳腺恶性的准确率和特异度高于GLR(79.2%对60.4%;87.5%对50.0%)。
  • 预压小于25%时FLR和GLR观察者间一致性均优(ICC 0.853和0.779)。

核心结论

  • FLR在检测乳腺恶性病变方面表现优于GLR,脂肪组织是计算恶性乳腺肿瘤弹性成像应变比的更佳参考组织;将预压保持在小于25%可使不同操作者获得可重复的FLR和GLR读数。

研究设计与样本

  • 对39例连续患者中基于超声发现而计划行空心针活检或切除手术的53个病灶行超声弹性成像,由2名超声医师独立在控制预压下采集,计算FLR和GLR并评估诊断性能与观察者间信度(ICC)。

诊断标准与关键数值

  • 以活检或手术为参考,FLR的准确率为79.2%、特异度为87.5%,GLR的准确率为60.4%、特异度为50.0%;恶性病灶的FLR和GLR均显著高于良性病灶(P < .05);预压小于25%时FLR的ICC为0.853(95%CI 0.738-0.920),GLR的ICC为0.779(95%CI 0.619-0.87)。

临床应用与阅片要点

  • 在乳腺弹性成像中采用脂肪组织作为参考组织计算应变比(FLR)可能较腺体参考(GLR)获得更高的准确率和特异度;采集时将预压控制在小于25%有助于提高不同操作者间FLR和GLR测量的可重复性。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The purpose of this study was to compare the diagnostic performance of the fat-to-lesion strain ratio (FLR) and gland-to-lesion strain ratio (GLR) for patients with indeterminate or suspicious breast lesions on ultrasound (US) imaging under a controlled precompression technique and to see whether the technique improves the reproducibility of FLR and GLR measurement.

METHODS: Fifty-three lesions in 39 consecutive patients who had scheduled core biopsy or excision surgery based on US findings were examined by US elastography. Each lesion was acquired under controlled precompression by 2 sonographers independently. Both the FLR and GLR of the lesion were calculated. For diagnostic performance, the sensitivity, specificity, negative predictive value, positive predictive value, and accuracy were obtained. Interobserver reliability between different sonographers was assessed by the intraclass correlation coefficient (ICC).

RESULTS: Forty lesions were benign, and 13 lesions were malignant. Both the FLR and GLR were significantly higher in malignant than benign lesions (P < .05). The FLR yielded higher accuracy and specificity compared to the GLR (accuracy, 79.2% versus 60.4%; and specificity, 87.5% versus 50.0%). With the controlled precompression applied at less than 25% during elastography, the interobserver agreement was excellent for FLR measurements (ICC, 0.853; 95% confidence interval, 0.738-0.920) and GLR measurements (ICC, 0.779; 95% confidence interval, 0.619-0.87).

CONCLUSIONS: The FLR performed better than the GLR in the detection of breast malignancy; thus, fatty tissue was a better reference tissue for calculating the strain ratio on malignant breast tumor elastography. Keeping precompression to less than 25% will enable different operators to acquire similar elastograms with reproducible FLR and GLR readings.

来源:J Ultrasound Med,查看原文。版权归原作者所有。

引用

  1. Chee C, Lombardo P, Schneider M, et al. Comparison of the Fat-to-Lesion Strain Ratio and the Gland-to-Lesion Strain Ratio With Controlled Precompression in Characterizing Indeterminate and Suspicious Breast Lesions on Ultrasound Imaging. J Ultrasound Med. 2019. PMID: 31140631. doi:10.1002/jum.15037

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。