单针电穿孔与间质电化学治疗:一种新系统的体内安全性与有效性评估
在10头家猪肝脏中评估单针多齿可膨胀电极施加不可逆电穿孔(IRE)与电化学治疗(IRECT)的安全性和有效性。技术成功9/10,1头于术中因室颤死亡;术后1天和3天增强CT显示IRECT消融体积显著大于单纯IRE。作者结论认为IRECT有效且消融体积更大,但心电图触发是安全应用的必要前提。
要点
- 术后1天增强CT示IRECT消融体积4.47±1.78 ml vs IRE 2.51±0.93 ml,差异显著(p<0.05)。
- 术后3天增强CT示IRECT消融体积3.39±1.05 ml vs IRE 1.53±0.78 ml,差异显著(p<0.05)。
- 技术成功9/10,1头猪术中因室颤死亡;作者指出心电图触发是安全应用的必要前提。
核心结论
- 使用该新系统的IRECT有效,且与单纯IRE相比提供显著更大的消融体积;心电图触发是允许该系统安全应用的必要前提。
研究设计与样本
- 体内动物实验,10头家猪;在CT引导下将单针多齿可膨胀电极施加器置于不同肝叶,一叶行常规IRE(1500 V,120个脉冲,脉冲长度100 μs),另一叶经可膨胀电极注射多柔比星混合物后行相同操作(IRECT);术后第1、3、7天行增强CT和MRI,相应在成像后第1、3、7天各处死3只动物并行组织病理学评估消融体积;采用相关t检验比较组间差异。
关键影像表现/征象
- 术后第1天和第3天随访CT显示IRECT与IRE消融体积差异显著(p<0.05),分别为4.47±1.78 ml vs 2.51±0.93 ml和3.39±1.05 ml vs 1.53±0.78 ml。
局限性
- 10头家猪中1头在干预期间因室颤死亡,技术成功为9/10。
- 作者结论指出心电图触发是允许该系统安全应用的必要前提。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: We conducted an in vivo trial to investigate the safety and efficacy of a newly developed system for the application of a combined therapy consisting of irreversible electroporation (IRE) and electrochemotherapy (IRECT) in the liver. The system is conceived as a single-needle multitined applicator with expandable electrodes that allow interstitial injection of fluids, e.g., chemotherapy.
METHODS: Experiments were conducted in ten domestic pigs. The applicator was placed in different liver lobes under CT guidance. In one lobe, the applicator was used for conventional IRE (1500 V, 120 pulses, pulse length 100 μs). In the other lobe, the same procedure was performed preceded by the injection of a doxorubicin mixture through the expandable electrodes (IRECT). Contrast-enhanced CT and MRI were performed on days 1, 3, and 7 after the procedure. Accordingly, three animals were sacrificed on days 1, 3, and 7 after the imaging and ablation volumes were evaluated histopathologically. Related t test was used to compare the groups.
RESULTS: Technical success was achieved in 9/10 experiments. One animal deceased during the intervention because of ventricular fibrillation. Follow-up CT 1 and 3 days after intervention showed a significant (p < 0.05) difference in the ablation volumes of IRECT vs IRE, respectively, of 4.47 ± 1.78 ml vs 2.51 ± 0.93 ml and of 3.39 ± 1.05 vs 1.53 ± 0.78 ml.
CONCLUSIONS: IRECT using the newly developed system proved to be effective and provided significantly larger ablation volumes compared with IRE alone. However, ECG triggering is a necessary prerequisite to allow a safe application of the system.
KEY POINTS: • Working on the geometry of the IRE applicator in terms of expandable electrodes may overcome the current limitations of IRE resulting from the placement of multiple electrodes. • Efficacy of IRE ablations can be enhanced by the interstitial application of chemotherapy in the periphery of ablation areas.