超声心动图混响伪影来源研究:体外实验与体内验证
研究提出心尖四腔心切面混响伪影产生的五种假说,通过体外实验与四名志愿者体内记录验证。结果显示混响伪影既可源于超声束路径内结构,也可源于成像平面外结构,肋骨镜面反射使声束偏转时平面外强回声结构可形成覆盖图像的混响伪影;不同组织来源的混响类型表现不同,结果有助于改进伪影抑制技术和探头设计。
要点
- 心尖四腔心切面混响伪影可源于成像平面外结构,肋骨镜面反射致声束偏转是重要机制。
- 不同组织来源的混响伪影在体内图像上表现不同,可据此判断来源。
- 结果有助于改进伪影抑制技术及设计高质量心脏超声探头。
核心结论
- 超声心动图混响伪影可起源于超声束路径内结构或成像平面外结构,肋骨镜面反射使声束偏转时平面外强回声结构可产生覆盖图像的混响伪影,不同组织来源的混响类型表现不同。
研究设计与样本
- 体外实验结合四名志愿者体内记录的对照研究,基于心尖四腔心切面提出五种混响伪影产生假说并验证其合理性。
关键影像表现/征象
- 成像平面外强回声结构的混响可在肋骨镜面反射使超声束偏转时于图像上形成混响伪影。
- 体内示例中不同混响类型外观不同,取决于其来源组织。
临床应用与阅片要点
- 结果可帮助心内科医生理解超声心动图混响伪影,并据伪影类型和来源获取更优图像。
- 结果可应用于改进混响抑制技术或设计高质量心脏图像超声探头。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Clutter in echocardiography hinders the visualization of the heart and reduces the diagnostic value of the images. The detailed mechanisms that generate clutter are, however, not well understood. We present five different hypotheses for generation of clutter based on reverberation artifact with a focus on apical four-chamber view echocardiograms. We demonstrate the plausibility of our hypotheses by in vitro experiments and by comparing the results with in vivo recordings from four volunteers. The results show that clutter in echocardiography can be originated both at structures that lie in the ultrasound beam path and at those that are outside the imaging plane. We show that reverberations from echogenic structures outside the imaging plane can make clutter over the image if the ultrasound beam gets deflected out of its intended path by specular reflection at the ribs. Different clutter types in the in vivo examples show that the appearance of clutter varies, depending on the tissue from which it originates. The results of this work can be applied to improve clutter reduction techniques or to design ultrasound transducers that give higher quality cardiac images. The results can also help cardiologists have a better understanding of clutter in echocardiograms and acquire better images based on the type and the source of the clutter.
引用
- Fatemi A, Berg EAR, Rodriguez-Molares A, et al. Studying the Origin of Reverberation Clutter in Echocardiography: In Vitro Experiments and In Vivo Demonstrations. Ultrasound Med Biol. 2019. PMID: 31053427. doi:10.1016/j.ultrasmedbio.2019.01.010
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。