心血管MRI金属伪影减少技术在心脏植入式电子装置患者中用于准确评估心肌瘢痕
本研究纳入50例心脏植入式电子装置(CIED)患者,在消融前采用常规与改良宽频反转恢复(IR)晚期钆增强(LGE)MRI技术,量化装置所致伪影体积并按轻、中、重三分位分层。常规LGE MRI在32/50(64%)病例中产生遮挡心室段的装置所致高信号伪影;宽频LGE MRI显著降低重度伪影体积(p<0.0001),完全消除所有轻度和多数中度伪影,总体伪影体积减少56%(p<0.0001)。
要点
- 常规LGE MRI在32/50(64%)CIED患者中产生遮挡心室段的高信号伪影。
- 宽频LGE MRI完全消除轻度和多数中度伪影,重度伪影体积显著降低(p<0.0001)。
- 宽频技术使队列总体伪影体积减少56%(p<0.0001)。
核心结论
- 改良宽频IR技术完全消除轻度与中度装置所致高信号伪影,并显著降低重度伪影体积,从而允许在消融前准确识别CIED患者的心肌瘢痕。
研究设计与样本
- 50例(17女、33男;平均年龄61±12岁;平均射血分数35.9%±13.3%)CIED患者在消融前接受常规与改良宽频IR LGE心脏MRI;量化装置所致伪影体积并按轻、中、重三分位分层;采用有序逻辑回归分析常规与宽频图像伪影体积的关联,并校正患者人口学特征。
关键影像表现/征象
- 常规LGE MRI在32/50(64%)病例中产生遮挡心室段的装置所致高信号伪影。
- 宽频LGE MRI完全消除所有轻度伪影和多数中度伪影。
- 宽频LGE MRI显著降低重度伪影体积(p<0.0001)。
- 常规序列最常遮挡段中持续存在伪影的病例数:基底前壁7/32、心室中部前壁8/32、心室中部前间隔4/32。
诊断标准与关键数值
- 装置所致伪影体积按三分位分层为轻度、中度、重度。
- 宽频LGE MRI使队列总体伪影体积减少56%(p<0.0001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE. An important application of late gadolinium enhancement (LGE) cardiac MRI is accurate assessment of myocardial scar before ablation. However, this is often limited in patients with cardiac implantable electronic devices (CIEDs) because of metal device-induced artifacts. The purpose of this study was to determine whether a modified wideband inversion recovery (IR) LGE MRI technique decreases artifact volume to allow the assessment of myocardial scar. SUBJECTS AND METHODS. Fifty patients (17 women and 33 men; mean age ± SD, 61 ± 12 years; mean ejection fraction ± SD, 35.9% ± 13.3%) with CIEDs underwent cardiac MRI using conventional and modified wideband IR LGE techniques before ablation. The volume of device-induced artifact was quantified and stratified by tertiles on mild, moderate, and severe. Ordinal logistic regression analysis assessed the association between artifact volume on conventional and wideband images adjusted for patients' demographics. RESULTS. Conventional LGE MRI resulted in device-induced hyperintense artifacts that obscured ventricular segments in 32 of 50 (64%) cases. Wideband LGE MRI significantly reduced severe artifact volume (p < 0.0001) and completely resolved all mild and most moderate artifacts. Overall, wideband techniques resulted in a 56% reduction in total artifact volume for the cohort (p < 0.0001). The wideband LGE MRI sequence minimized artifacts in the most commonly obscured segments on the conventional LGE MRI sequence, with persistent artifacts in seven, eight, and four of 32 cases at the basal anterior, midventricular anterior, and midventricular anteroseptal segments, respectively. CONCLUSION. The modified wideband IR technique completely resolves mild and moderate device-induced hyperintense artifacts and significantly reduces the volume of severe artifact to allow accurate identification of myocardial scar in patients with CIEDs before ablation.