基于风险分析方法预防透视引导介入操作造成的伤害
研究针对美国退伍军人事务部(VA)及非VA医疗机构中透视引导介入操作可能带来的高辐射剂量风险,采用医疗失效模式与效应分析(HFMEA)对植入式心律转复除颤器导线拔除术进行示例分析,识别并缓解系统性缺口。共制定29项干预措施,并按可行性、成本效益和预期临床影响排序,其中5项被推荐立即实施或试点。作者认为HFMEA等正式风险分析技术可推荐用于其他机构的高风险操作安全管理。
要点
- 峰值皮肤剂量超过3 Gy的患者可能发生放射性组织损伤,从短暂红斑到不愈合伤口。
- 辐射损伤可在术后数周至数月才显现,导致正确诊断和及时医疗干预更困难。
- 研究通过HFMEA识别系统性缺口,29项干预中5项被推荐立即实施或试点。
核心结论
- 本研究展示了正式风险分析技术在透视引导介入操作中的应用;高优先级干预措施可能对其他机构在实施潜在高辐射剂量操作和进行风险-获益分析时有参考价值;推荐其他机构使用HFMEA等正式风险分析技术改善高风险操作的安全性。
研究设计与样本
- 多机构团队运用医疗失效模式与效应分析(HFMEA)对一例植入式心律转复除颤器导线拔除术进行示例分析,研究涉及美国退伍军人事务部(VA)及非VA医疗机构。
临床应用与阅片要点
- 作者推荐其他机构使用HFMEA等正式风险分析技术改善高风险操作的安全性;29项干预措施中有5项被推荐立即实施或试点。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: Fluoroscopically guided interventional (FGI) procedures often have lower complication rates compared with alternative surgical procedures, providing an option for patients with a high risk of perioperative mortality. Although severe radiation injuries are rare, patients receiving peak skin doses exceeding 3 Gy can suffer from radiation-induced tissue injuries, ranging from transient erythema to nonhealing wounds. As these iatrogenic injuries may manifest weeks to months postprocedure, proper diagnosis and timely medical intervention are less likely. Clinically, the lack of situational awareness for monitoring air kerma continues to be a challenge despite the recommendations of numerous organizations on ways to achieve fluoroscopy safety. For patient safety efforts, this study aimed to identify and mitigate systematic gaps associated with potentially high-radiation dose fluoroscopic procedures in US Department of Veterans Affairs (VA) and non-VA medical institutions.
METHODS: In this study, a multi-institutional team utilized Healthcare Failure Mode and Effect Analysis (HFMEA) on an example implantable cardioverter defibrillator lead extraction procedure.
RESULTS: With this approach, 29 interventions were devised and prioritized by feasibility, cost-effectiveness, and expected clinical impact. Five of the 29 interventions were recommended for immediate implementation or piloting.
CONCLUSIONS: This work demonstrates the application of formal risk-based analysis techniques in FGI procedures. These high priority interventions may be valuable for other facilities to consider when performing potentially high-radiation dose procedures and conducting risk-benefit analyses. Formal risk analysis techniques such as the HFMEA process are recommended for other facilities to use to improve safety for their high-risk procedures.
引用
- Justinvil GN, Leidholdt EM, Balter S, et al. Preventing Harm From Fluoroscopically Guided Interventional Procedures With a Risk-Based Analysis Approach. J Am Coll Radiol. 2019. PMID: 31031169. doi:10.1016/j.jacr.2019.02.047
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。