患者体型与辐射剂量对双层探测器CT碘定量及虚拟平扫值准确性和精密度的影响:体模研究

Impact of Patient Size and Radiation Dose on Accuracy and Precision of Iodine Quantification and Virtual Noncontrast Values in Dual-layer Detector CT-A Phantom Study.
PubMed 文献 · Acad Radiol · 2019-04-12

该体模研究用双层探测器CT扫描三种模拟儿童、平均及大体重成人的体模,碘浓度0–32 mg/ml,CTDIvol 2–15 mGy,小体模另扫0.9–1.8 mGy。结果显示碘定量和虚拟平扫的准确性主要取决于体模大小而非剂量,低剂量儿童方案对碘定量和虚拟平扫衰减值有显著但有限的影响。

要点

  • 碘定量与虚拟平扫的准确性主要受患者体型影响,而非辐射剂量。
  • 在2–15 mGy范围内,各体模大小的平均虚拟平扫衰减和碘定量准确性无显著差异。
  • 小体模在2 mGy和10 mGy时碘定量精密度显著降低。

核心结论

  • 双层探测器CT的碘定量和虚拟平扫减影性能在很大程度上与剂量无关,主要影响因素是患者体型;低剂量儿童扫描方案对碘定量和虚拟平扫衰减值有显著但有限的影响。

研究设计与样本

  • 体模研究,采用双层探测器CT扫描三种模拟10岁儿童、平均体型成人和大体重成人的体模,碘溶液插入物浓度范围0–32 mg/ml,每种体模在120和140 kVp下以CTDIvol 2–15 mGy扫描,最小体模额外以CTDIvol 0.9–1.8 mGy扫描,所有扫描重复3次。

诊断标准与关键数值

  • 在2–15 mGy扫描范围内,小、中、大体模碘插入物的平均虚拟平扫衰减误差分别为1.2 HU ± 3.2、-1.2 HU ± 14.9、2.6 HU ± 23.6。
  • 在2–15 mGy扫描范围内,小、中、大体模碘插入物的碘定量误差分别为+0.1 mg/cc ± 0.4、-0.9 mg/cc ± 0.9、-1.8 mg/cc ± 1.8。
  • 在该剂量范围内,各体模大小的平均虚拟平扫衰减或碘定量准确性无显著差异(p ≥ 0.05),但小体模在2 mGy和10 mGy时碘定量精密度显著降低(p < 0.05)。
  • 小体模以CTDIvol 0.9–1.8 mGy扫描时,虚拟平扫衰减精密度和碘定量准确性分别为-0.5 HU ± 5.3和-0.3 mg/cc ± 0.5,较同体模更高剂量扫描显著降低但程度有限。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: Iodine quantification (IQ) and virtual noncontrast (VNC) images produced by dual-energy CT (DECT) can be used for various clinical applications. We investigate the performance of dual-layer DECT (DLDECT) in different phantom sizes and varying radiation doses and tube voltages, including a low-dose pediatric setting.

MATERIALS AND METHODS: Three phantom sizes (simulating a 10-year-old child, an average, and a large-sized adult) were scanned with iodine solution inserts with concentrations ranging 0-32 mg/ml, using the DLDECT. Each phantom size was scanned with CTDIvol 2-15 mGy at 120 and 140 kVp. The smallest phantom underwent additional scans with CTDIvol 0.9-1.8 mGy. All scans were repeated 3 times. Each iodine insert was analyzed using VNC and IQ images for accuracy and precision, by comparison to known values.

RESULTS: For scans from 2 to 15 mGy mean VNC attenuation and IQ error in the iodine inserts in the small, medium, and large phantoms was 1.2 HU ± 3.2, -1.2 HU ± 14.9, 2.6 HU ± 23.6; and +0.1 mg/cc ± 0.4, -0.9 mg/cc ± 0.9, and -1.8 mg/cc ± 1.8, respectively. In this dose range, there were no significant differences (p ≥ 0.05) in mean VNC attenuation or IQ accuracy in each phantom size, while IQ was significantly less precise in the small phantom at 2 mGy and 10 mGy (p < 0.05). Scans with CTDIvol 0.9-1.8 mGy in the small phantom showed a limited, but statistically significantly lower VNC attenuation precision and IQ accuracy (-0.5 HU ± 5.3 and -0.3 mg/cc ± 0.5, respectively) compared to higher dose scans in the same phantom size.

CONCLUSION: Performance of iodine quantification and subtraction by VNC images in DLDECT is largely dose independent, with the primary factor being patient size. Low-dose pediatric scan protocols have a significant, but limited impact on IQ and VNC attenuation values.

来源:Acad Radiol,查看原文。版权归原作者所有。